Multi-Laboratory Assessment Reveals Variable Ion Species Profiles in Electrospray Ionization Mass Spectrometry
Melanie T. Odenkirk, Jacqueline M. Chaparro, Nathan T. Montgomery, Katrina L. Leaptrot, Stacy D. Sherrod, Jody C. May, Henriëtte van Eekelen, Bert Schipper,
Colorado State University · Vanderbilt University
Key points
From the paper's conclusion- These findings demonstrate that, even with standardized methods, the collective influence of local ionization environments on ion species formation remains a critical obstacle for interpreting LC-MS small molecule data and improving reproducibility and comparability across studies.
Abstract
Electrospray ionization (ESI) is a foundational technique in mass spectrometry (MS) widely applied to measure nonvolatile molecules in diverse chemical and biological samples. Yet variation in ESI-MS instrumentation, operating settings, and parameters such as solvent composition creates specific, local ionization environments that drive the formation of select ion species for individual analytes. As ESI continues to advance analytical discovery, understanding the extent to which variation in ion species formation impacts intra- and inter-experimental results is essential. Here, we assessed ion species formation by analyzing an internal retention time standard (IRTS) mixture across ten laboratories employing high-resolution ESI-MS instrumentation from four vendors (Agilent Technologies, ThermoFisher Scientific, Shimadzu Corporation, and Waters Corporation). Instrument vendors were considered not as a benchmark of performance, but as a practical framework to capture differences in source design, ion optics, and analyzer/detectors that are inherently coupled to commercial platforms. Despite the use of standardized extraction and chromatographic protocols, differences in instrument configuration, source conditions, and method execution resulted in variation in ion species formation across vendors, among laboratories using instruments from the same vendor, and even within individual laboratories. These findings demonstrate that, even with standardized methods, the collective influence of local ionization environments on ion species formation remains a critical obstacle for interpreting LC-MS small molecule data and improving reproducibility and comparability across studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance comparison of UniSprayTM and electrospray ionisation coupled to ion-mobility high-resolution mass spectrometer for screening of organic pollutants in water matrices.Journal of chromatography. A · 2026
- Electrospray Regime Controls Sensitivity and Quantitative Performance in Micro- and Nano-Flow CE–ESI–MSAnalytical Chemistry · 2026
- An Energetic-Control Framework for Electrospray Ionization Using Thermometer Ions.Journal of the American Society for Mass Spectrometry · 2026
- AGENTi: APCI-ESI Analyte Ionization Fidelity Pre-Screening Workflow using Multimodel Inference.Journal of chemical information and modeling · 2026
- From Innovation to Translation: Emerging Direct‐Analysis Mass Spectrometry PlatformsJournal of Mass Spectrometry · 2026
- Titration-Based Assessment of Matrix Effects in Direct Infusion Mass Spectrometry.Journal of the American Society for Mass Spectrometry · 2026