Radiation dose perturbations associated with plastic and metallic tracheostomy tubes: a phantom study.
Tsung-Yu Yen, Ke-Yu Lien, Hsiao-Mei Fu, Chien-Fu Lin, Jehn-Chuan Lee, Shih-Ming Hsu
MacKay Memorial Hospital, Department of Radiation Oncology, Taipei, Taiwan; MacKay Memorial Hospital, Hospice and Palliative Care Center, Taipei, Taiwan; MacKay Medical University, Department of Medicine, Taipei, Taiwan; MacKay Junior College of Nursing, Medicine and Management, Department of Death Care Service, Taipei, Taiwan; National Yang Ming Chiao Tung University, Department of Biomedical Imaging and Radiological Sciences, Taipei, Taiwan. · National Yang Ming Chiao Tung University, Department of Biomedical Imaging and Radiological Sciences, Taipei, Taiwan; Koo Foundation Sun Yat-Sen Cancer Center, Department of Radiation Oncology, Taipei, Taiwan.
Key points
From the paper's conclusion- Metallic tracheostomy tubes induce substantial mucosal dose perturbations underestimated by the AAA algorithm, indicating they are ill-suited for radiotherapy. When clinically feasible, substituting metallic tubes with plastic alternatives is recommended prior to irradiation. If exchange is contraindicated, clinicians must remain vigilant regarding unmodeled mucosal hyperdosage.
Abstract
OBJECTIVE: To evaluate mucosal dose perturbations from Polyvinyl Chloride (PVC) and metallic tracheostomy tubes during radiotherapy, and Treatment Planning System (TPS) calculation accuracy against Thermoluminescent Dosimeter (TLD) measurements. METHODS: A customized head-and-neck PMMA phantom evaluated mucosal doses across three conditions (no tube, PVC tube, metallic tube) using 6-MV photon beams under Anterior-Posterior (AP), Parallel-opposed (AP‒PA), and central-block configurations. Doses were measured at five paired points above (A1-E1) and below (A2-E2) the tube using ultrathin TLDs and compared with TPS (Eclipse AAA algorithm) calculations. Pairwise comparisons utilized paired Student'st-tests with Bonferroni correction (padj < 0.05). RESULTS: TLD measurements revealed marked dose escalation with metallic tubes. Under AP delivery, upper mucosal dose was significantly higher for metallic than PVC tubes (227.82 ± 11.45 vs. 203.04 ± 8.17 cGy; padj < 0.01). Under AP/PA delivery, metallic tubes induced substantial bilateral dose enhancements on upper (258.12 ± 9.68 vs. 204.68 ± 6.94 cGy; padj < 0.01; peak +38.79%) and lower surfaces (255.66 ± 20.31 vs. 199.82 ± 5.15 cGy; padj < 0.05). Central block delivery effectively suppressed surface doses (<12 cGy) without differences. The TPS with AAA algorithm predicted deterministic variations within only 1%-3%, severely underestimating the empirical interface hyperdosage. CONCLUSION: Metallic tracheostomy tubes induce substantial mucosal dose perturbations underestimated by the AAA algorithm, indicating they are ill-suited for radiotherapy. When clinically feasible, substituting metallic tubes with plastic alternatives is recommended prior to irradiation. If exchange is contraindicated, clinicians must remain vigilant regarding unmodeled mucosal hyperdosage.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of wedge-induced photoneutron dose during 15 MV 3D conformal rectal radiotherapy using TLD600/TLD700 pairs.Journal of X-ray science and technology · 2026
- Dosimetric impact of titanium, zirconia, and customized subperiosteal dental implants in head and neck radiotherapy: an in vitro study.Frontiers in oncology · 2026
- Evaluation of radiation-protection drapes for reducing operator and patient radiation dose during overtable fluoroscopy: a phantom study.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Spatial distribution of X-ray attenuation in a full-face shield: a multipoint in vivo study in interventional cardiology.Radiation protection dosimetry · 2026
- Occupational exposure assessment in ROLL procedures with Tc-99m: A Monte Carlo simulation approach.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Relative transmitted-dose reduction of 3D-printed and elastomeric dental radiotherapy spacers at different thicknesses.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026