Relative transmitted-dose reduction of 3D-printed and elastomeric dental radiotherapy spacers at different thicknesses.
Wisarut Prawatvatchara, Tarnsai Tantiniramai, Thireeya Apasrawirote, Apinya Limvisitsakul, Pipop Saikaew, Cheewin Towithelertkul, Dusit Nantanapiboon
Department of Prosthodontics, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand. · Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand.
Key points
From the paper's conclusion- Under the tested phantom and irradiation conditions, material and thickness influenced the relative transmitted-dose reduction. The highest transmitted dose reduction was observed in PVS-P. However, the effects were modest and thickness- dependent. These setup-specific effects do not establish intrinsic material attenuation or clinical tissue protection.
Abstract
PURPOSE: To evaluate the radiation dose reduction of different materials for intraoral radiotherapy prostheses and spacers with different thicknesses. MATERIALS AND METHODS: Three additively manufactured resins-KeySplint Hard (3D-Hard), Temporary CB Resin (3D-TCB), and Saremco Print CROWNTEC (3D-PCB)-and three elastomeric materials-putty-type polyvinyl siloxane (PVS-P), light-body polyvinyl siloxane (PVS-L), and polyether (PE)-were prepared with two thicknesses (5 and 10 mm, n = 5 per group). Rectangular specimens were fabricated and positioned over an ionization chamber in a solid-water phantom and measured 15 times with a 6 × 8 cm2 field. The relative transmitted dose reduction was calculated against the bracketed no-specimen reference values. Statistical analysis was done using two-way analysis of variance followed by Games-Howell multiple comparisons. RESULTS: Material and thickness significantly affected the relative transmitted dose reduction (p < 0.05), whereas their interaction was not significant (p > 0.05). PVS-P had significantly higher transmitted dose reduction than other materials, regardless of material thickness (3.50% ± 0.21% at 5 mm and 5.57% ± 0.43% at 10 mm). Among the additively manufactured resins, 3D-PCB demonstrated higher transmitted dose reduction, where a significant effect was found at 10 mm thickness. At 5 mm, no significant difference was observed among three additively manufactured resins. However, at 10 mm, 3D-PCB (3.83% ± 0.98%) produced significantly greater reduction than 3D-Hard (3.06% ± 1.05%) and 3D-TCB (3.13% ± 0.98%). For material thickness, the 10-mm specimens produced greater reduction than the 5-mm specimens. CONCLUSIONS: Under the tested phantom and irradiation conditions, material and thickness influenced the relative transmitted-dose reduction. The highest transmitted dose reduction was observed in PVS-P. However, the effects were modest and thickness- dependent. These setup-specific effects do not establish intrinsic material attenuation or clinical tissue protection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Study on the influence of fixed dental prostheses on radiotherapy dose distribution based on dual-layer detector spectral CT imaging.Frontiers in oncology · 2026
- ABS and PETG as candidate materials for 3D-printed mammography phantoms: Physical and radiological characterization and assessment of TLD-indicated air kerma as a function of depth.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Comparison of three dental resins for 3D printing of orthodontic appliances: Comparison of leaching, biocompatibility, and thermo-mechanical properties after post-curing and agingDental Materials · 2025
- Comparative Evaluation of the Effect of Radiation on the Marginal Adaptation of Different Composite Restorative Materials: An In Vitro Study.Cureus · 2026
- Dosimetric impact of titanium, zirconia, and customized subperiosteal dental implants in head and neck radiotherapy: an in vitro study.Frontiers in oncology · 2026
- Effect of fractionated therapeutic radiotherapy on surface microhardness and roughness of giomer- and ormocer-based restorative composites: an in vitro studyScientific Reports · 2026