Unexpected Metal Ion Elevation and Adverse Tissue Reactions in Trident II Dual Mobility Constructs With Ceramic Heads.
Henry Somerby, Carmelo Burgio, Gabrielle Dykhouse, Edward Grabov, Aarti Shenoy, Daniel Ramirez, Jonathan Vigdorchik, Gwo-Chin Lee,
Stavros Niarchos Foundation Complex Joint Reconstruction Center, Hospital for Special Surgery, New York, NY, USA. · Department of Orthopaedic Surgery, Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, New York, NY, USA.
Key points
From the paper's conclusion- Surviellance through serum metal ion testing and magnetic resonance imaging may prevent soft tissue damage and facilitate timely revision.
Abstract
Dual mobility (DM) constructs reduce dislocation risk in total hip arthroplasty. The Trident II system accepts a cobalt-chromium liner to create a DM bearing. However, dissimilar metal at the cup-liner interface raises concerns for corrosion. This case series describes six cementless primary Trident II DM total hip arthroplasty patients who experienced gradual onset of hip pain with low-grade inflammatory symptoms and elevated serum cobalt. All patients underwent revision surgery for metallic debris reaction. Preoperative magnetic resonance imaging revealed local tissue reaction, and intraoperative findings confirmed corrosion at the cup-liner interface. Surviellance through serum metal ion testing and magnetic resonance imaging may prevent soft tissue damage and facilitate timely revision. As DM systems evolve, further investigation is warranted to evaluate safety.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Is It the Trunnion or Is It the Liner? Investigating Bilateral Dual-Mobility Constructs With Different Inner Femoral Head Compositions.Arthroplasty today · 2026
- Intraprosthetic dislocation in modular dual-mobility: does the company matter? A systematic review and meta-analysis.Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2026
- Corrosion and wear damage estimations for retrieved knee and hip implants.Npj Materials degradation · 2026
- Total hip arthroplasty with dual-mobility components is associated with excellent clinical outcomes and low metal ion blood levels at 2 years.Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2026
- Metallosis Following Bipolar Hemiarthroplasty Due to Polyethylene Wear: A Case Report.Cureus · 2026
- Evaluating the Potential Cardiac Toxicity of Implant-Derived Metal Ions and Particles Using H9C2 and Stem Cell Model.ACS applied materials & interfaces · 2026