Metallosis Following Bipolar Hemiarthroplasty Due to Polyethylene Wear: A Case Report.
Gangadhar Gundu, Aditya Krishna Mootha, Uma Shankar Bajjangi, Umamahesh Neeli, Venkat Mallipudi
Orthopaedics and Traumatology, Siddhartha Hospital, Kakinada, IND. · Arthroplasty, Siddhartha Hospital, Kakinada, IND.
Key points
From the paper's conclusion- We report the case of a 54-year-old man who presented with six months of progressive right hip pain, four years after bipolar hemiarthroplasty for a transcervical fracture of the femoral neck.
Abstract
Metallosis following bipolar hemiarthroplasty is an uncommon complication that can result from wear of the concealed polyethylene bearing within the bipolar head. We report the case of a 54-year-old man who presented with six months of progressive right hip pain, four years after bipolar hemiarthroplasty for a transcervical fracture of the femoral neck. Laboratory investigations showed no evidence of infection. Radiographs and computed tomography demonstrated eccentric positioning of the femoral head within the bipolar cup, with peri-trochanteric and acetabular osteolysis and a contained supra-acetabular bone defect measuring approximately 2 × 2 cm. During revision surgery, extensive blackish discoloration of the periprosthetic tissues consistent with metallosis was observed. Examination of the retrieved bipolar head demonstrated complete wear-through of the inner polyethylene liner, resulting in direct metal-on-metal articulation between the inner and outer metal components and generation of substantial metallic debris. A two-stage revision was performed. Extended trochanteric osteotomy was required for femoral stem extraction, followed by extensive debridement and placement of an antibiotic cement spacer. After eight weeks, iliac crest bone grafting was performed for the acetabular defect and the patient underwent conversion to cementless total hip arthroplasty. Histopathological examination confirmed metallosis. At follow-up, the patient was pain-free and independently mobile. This case highlights that concealed polyethylene wear within a bipolar head can progress to direct metal-on-metal articulation, metallosis, osteolysis, and prosthetic failure. Eccentric positioning of the femoral head in a painful bipolar hemiarthroplasty should prompt careful evaluation for polyethylene bearing failure, even when inflammatory markers are normal. Early recognition and appropriate staged revision can facilitate management of extensive metallosis and bone loss.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Metallosis Following Total Hip Arthroplasty: A Case Report.Journal of orthopaedic case reports · 2026
- Revision total hip arthroplasty for osteolysis and metallosis 17 years post-THA: a case report and literature review.International journal of surgery case reports · 2026
- Is It the Trunnion or Is It the Liner? Investigating Bilateral Dual-Mobility Constructs With Different Inner Femoral Head Compositions.Arthroplasty today · 2026
- Transcutaneous extrusion of a migrated bipolar hip hemiarthroplasty 9 years after implantation: a case report.Journal of surgical case reports · 2026
- Catastrophic Failure of a Prosthetic Femoral Stem due to Component Malposition: A Case Report.Case reports in orthopedics · 2026
- Catastrophic Acetabular Osteolysis and Component Migration Secondary to Surgical Cup Malpositioning: A Case Report.Cureus · 2026