Intraprosthetic dislocation in modular dual-mobility: does the company matter? A systematic review and meta-analysis.
Jean Shanaa, Jacob Stoebner, Theodor Di Pauli von Treuheim, Ran Schwarzkopf, Anna Cohen-Rosenblum, Vinay K Aggarwal
California Northstate University College of Medicine, Elk Grove, CA, USA. · Department of Orthopaedic Surgery, NYU Langone Health, New York, NY, USA.
Key points
From the paper's conclusion- The incidence of IPD in MDM implants is low, with pooled rates remaining below 1%. Revision rates were modest and consistent with published data on complex hip arthroplasty. No significant differences between manufacturers were detected, though findings are limited by the small number of non-Stryker implants reported. These findings suggest that MDM designs appear safe with overall low numbers of IPD-related complications, however, larger registry-based studies are needed to better define potential design-related differences in IPD risk.
Abstract
INTRODUCTION: Intraprosthetic dislocation (IPD) is a rare but potentially catastrophic complication of dual-mobility (DM) implants, defined by dissociation of the inner femoral head from the polyethylene liner. Modular dual-mobility (MDM) designs were introduced to improve intraoperative flexibility and expand indications for DM technology. However, increasing reports of IPD have raised concerns that implant design and manufacturing differences may influence risk. The true incidence of IPD across MDM manufacturers remains unclear. METHODS: A systematic search of PubMed, Embase, and Web of Science was performed. Studies reporting IPD incidence and stratified outcomes by implant manufacturer were included. A random-effects proportion meta-analysis was conducted to estimate pooled IPD incidence and revision rates overall and by manufacturer. RESULTS: Of 523 identified articles, 32 met inclusion criteria, representing 3996 implants (2711 primary; 1285 revision). Mean follow-up was 3.48 years and mean age was 67.48 years. The overall IPD incidence was 0.38% (15/3996), with a pooled estimate of 0.1% (95% CI, 0-0.5%). Stratified analyses showed IPD rates of 0% for Stryker (95% CI, 0-0.7%), 0% for Zimmer (95% CI, 0-100%), and 0% for Groupe Lépine (95% CI, 0-100%). Reoperation rates were 1.4% (95% CI, 0.9-2.1%) in primary and 8.3% (95% CI, 6.0-11.3%) in revision THA. CONCLUSIONS: The incidence of IPD in MDM implants is low, with pooled rates remaining below 1%. Revision rates were modest and consistent with published data on complex hip arthroplasty. No significant differences between manufacturers were detected, though findings are limited by the small number of non-Stryker implants reported. These findings suggest that MDM designs appear safe with overall low numbers of IPD-related complications, however, larger registry-based studies are needed to better define potential design-related differences in IPD risk.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Total hip arthroplasty with dual-mobility components is associated with excellent clinical outcomes and low metal ion blood levels at 2 years.Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2026
- Single versus dual mobility trapeziometacarpal prostheses: Any short-term differences?Annales de chirurgie plastique et esthetique · 2026
- 10-Year Follow-Up Outcomes of a Modular Dual-Mobility in Primary Total Hip Arthroplasty.The Journal of arthroplasty · 2026
- Unexpected Metal Ion Elevation and Adverse Tissue Reactions in Trident II Dual Mobility Constructs With Ceramic Heads.Arthroplasty today · 2026
- Statistical Analysis Plan for the REDEP study: Effectiveness of dual-mobility cups for preventing dislocation after primary total hip arthroplasty by a posterolateral approach compared to unipolar cups in elderly patientsmedRxiv · 2026
- The 10-Year Revision Risk of Dual-Mobility Versus Standard Cups in Primary Total Hip Arthroplasty for Osteoarthritis: A Population-Based Study from the Danish Hip Arthroplasty Register.The Journal of arthroplasty · 2026