Corrosion and wear damage estimations for retrieved knee and hip implants.
Saman Nikpour, Anastasia M Codirenzi, Christine Hoskin, Ubong Eduok, Brent Lanting, Matthew G Teeter, Camila P E de Souza, Yolanda S Hedberg
Department of Chemistry, The University of Western Ontario, London, ON Canada. · Department of Medical Biophysics, Schulich School of Medicine and Dentistry, The University of Western Ontario, London, ON Canada.
Key points
From the paper's conclusion- This study scored mechanical and corrosion damage based on optical microscopy (OM) for 135 trunnion parts of retrieved hip implants and 106 tibial baseplates of retrieved knee implants.
Abstract
Predicting the lifetime and function of orthopedic implants requires learning from failures observed in retrieved implants. This study scored mechanical and corrosion damage based on optical microscopy (OM) for 135 trunnion parts of retrieved hip implants and 106 tibial baseplates of retrieved knee implants. We used scanning electron microscopy (SEM) and energy dispersive X-ray spectroscopy to categorize the damage types. This data was correlated with implant and patient factors. Fretting, third-body wear, tribocorrosion, and biofilms were observed. Mechanically assisted corrosion, e.g., tribocorrosion, was the dominant damage mode for both hip and knee implants. The knee implant damage strongly depended on the locking mechanism of the polyethylene insert on the tibial baseplate. For the trunnion part, infection prior to or at surgery was correlated with higher SEM damage scores. Inflammatory arthritis and cemented hip implants correlated with lower SEM damage scores of the trunnion. Damage scores decreased with patient age, and increased with patient weight, body mass index, and years of implantation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Tribocorrosion of Metallic Biomaterials under Simulated Inflammation Conditions: Current Status and Future ProspectsAdvanced Healthcare Materials · 2025
- Machine learning-based detection and area quantification of mechanical and corrosion damage in femoral stem tapers.Journal of the mechanical behavior of biomedical materials · 2026
- Biotribology of meniscus degeneration: correlations between frictional, structural and compressive properties.Frontiers in bioengineering and biotechnology · 2026
- Unexpected Metal Ion Elevation and Adverse Tissue Reactions in Trident II Dual Mobility Constructs With Ceramic Heads.Arthroplasty today · 2026
- Linking Circulating Metal Levels to Cytokine Signatures in Total Joint Arthroplasty Patients.ACS biomaterials science & engineering · 2026
- Prosthesis-Related Risk Factors for Trunnionosis: A Comprehensive Literature Review.Clinics in orthopedic surgery · 2026