Analytical performance of plastic scintillation microspheres versus a liquid scintillation cocktail for measuring beta-emitting radionuclides in representative aqueous media.

Hui Zhang, Yan Ma, Lina Ma, Xuewen Yan, Ying Lu

China Institute for Radiation Protection, Taiyuan, 030006, Shanxi, China; Shanxi Key Laboratory of Nuclear Medicine Research and Translation and Precision Protection, Taiyuan, 030006, Shanxi, China. Electronic address: zhanghui@cirp.org.cn. · China Institute for Radiation Protection, Taiyuan, 030006, Shanxi, China; Shanxi Key Laboratory of Nuclear Medicine Research and Translation and Precision Protection, Taiyuan, 030006, Shanxi, China.

Key points

From the paper's conclusion
  • Polystyrene-based plastic scintillation microspheres (PSm) doped with PPO and bis-MSB were therefore evaluated as a heterogeneous counting medium and compared with OptiPhase HiSafe 3 in representative aqueous matrices.

Abstract

Liquid scintillation counting provides high efficiency for beta emitters, but strongly acidic, alkaline, and saline samples can restrict sample loading and increase blank signals. Polystyrene-based plastic scintillation microspheres (PSm) doped with PPO and bis-MSB were therefore evaluated as a heterogeneous counting medium and compared with OptiPhase HiSafe 3 in representative aqueous matrices. The prepared PSm were spherical, with particle sizes mainly of 110-230 μm. Their response was strongly energy dependent: counting efficiencies were 0.38% for 3H, 40.2% for 14C, and an apparent 184.6% for 90Sr/90Y at secular equilibrium when normalized to 90Sr activity, compared with 41.5%, 93.2%, and 198.9%, respectively, for HiSafe 3. PSm allowed a fixed sample loading ratio of 0.50 mL/g under all tested matrix conditions and exhibited blank count rates of 20.9-28.1 cpm when measured for 10 min after 24 h of dark storage. The operational loading capacity of HiSafe 3 varied from 0.04 to 0.80 mL/mL. Following approximately 24 h of dark storage, its blank count rate reached 246.6 cpm in 4 mol/L NaOH during a 10 min measurement. For 14C, under the standardized conditions of 24 h dark storage and 10 min counting, PSm gave calculated minimum detectable activity concentrations (MDCs) of 138.8-170.2 Bq/L. HiSafe 3 gave lower MDCs for water, 1 mol/L NaOH, and 1 mol/L NaCl, whereas PSm gave lower MDCs in 2-4 mol/L NaOH, all tested HCl matrices, and 2-4 mol/L NaCl, including a 6.8-fold lower MDC in 2 mol/L HCl. Independent quantitative evaluation of the PSm system was limited to water and 1 mol/L HCl, NaOH, and NaCl over 1.8-200 kBq/L, with relative deviations within 11.1%. The lower calculated MDCs indicate potential detection advantages in concentrated matrices, where quantitative performance remains to be validated.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers