Possible underestimation in surface contamination monitoring: Radionuclide-specific swab selection for 3H and 14C.
Ciotoli Alessandro, Ampollini Marco, Maiorana Andrea, Di Carlo Christian
Sapienza - University of Rome, Department of Electrical and Energetic Engineering, Via Eudossiana, Rome 18 - 00184, Italy. · Italian National Institute of Health, National Center for Radiation Protection and Computational Physics, Viale Regina Elena, Rome 299 - 00161, Italy.
Key points
From the paper's conclusion- These findings demonstrate that reliable smear tests can be performed with standard laboratory consumables, without the need for specialised products, provided that the swab material is selected on a radionuclide-specific basis.
Abstract
The monitoring of surface contamination is a key element of radiological protection programmes in facilities handling unsealed radioactive sources, such as biological, clinical and radiochemical laboratories. For low-energy β-emitters such as tritium (³H) and carbon-14 (¹⁴C), direct detection is generally ineffective, and smear sampling followed by liquid scintillation counting (LSC) represents the standard approach. However, the swab material may affect the counting process and lead to activity underestimation, making the choice of a suitable substrate a relevant issue for routine monitoring. In the present work, five readily available laboratory materials were comparatively assessed as swab substrates: polystyrene, glass fibre, nylon, polycarbonate and polytetrafluoroethylene (PTFE). To isolate the effect of the substrate from sampling uncertainties, known activities of ³H and ¹ ⁴C were deposited directly onto each material and measured using the Triple-to-Double Coincidence Ratio (TDCR) method. Polystyrene, which fully dissolves in the scintillation cocktail, yielded recoveries close to 100% for both radionuclides, although its poor mechanical resistance limits its use for wiping. Glass fibre, nylon and PTFE also showed recoveries close to 100% for ³H, whereas for ¹ ⁴C geometric shielding reduced recovery to approximately 70% and 50% for unfolded and folded filters, respectively. Polycarbonate caused colour quenching, with ³H losses of 75-100%. These findings demonstrate that reliable smear tests can be performed with standard laboratory consumables, without the need for specialised products, provided that the swab material is selected on a radionuclide-specific basis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Analytical performance of plastic scintillation microspheres versus a liquid scintillation cocktail for measuring beta-emitting radionuclides in representative aqueous media.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Evaluation of Contamination Meter Efficiency for Detecting Surface Activity Across Decay Modes and Contaminated Area Sizes in Clinical Settings.Health physics · 2026
- Electrolytic enrichment and liquid scintillation determination of low-level tritium in high-salinity seawater with correction for salinity matrix effectsRadiochimica Acta · 2026
- Matrix effects on gross alpha determination in environmental water residues: an analytical framework based on effective alpha-particle mass range.Environmental geochemistry and health · 2026
- Engineered Activated Carbons for Multi-Radionuclide Removal from Real Low-Level Radioactive Wastewater.Nanomaterials (Basel, Switzerland) · 2026
- Irradiated graphite decontamination: 14C and 3H release–a steam-assisted thermal treatment approachJournal of Radioanalytical and Nuclear Chemistry · 2026