Numerically guided development of a microfluidic lab-on-chip platform for Cherenkov detection of β-emitting radionuclides.
Sarah E Lu, Geraldine Clinton-Bailey, Frances Burrell, Richard I Marsh, Jake Ludgate, Christopher L Cardwell, Reuben Forrester, Andrew Morris,
University of Southampton, National Oceanography Centre, European Way, Southampton, SO14 3ZH, UK. Electronic address: sarah.lu@soton.ac.uk. · Ocean Technology and Engineering Group, National Oceanography Centre, Southampton, SO14 3ZH, UK.
Key points
From the paper's conclusion- This work demonstrates the value of modelling-guided optimisation for the development of radiometric microfluidic analytical platforms and highlights the potential of Cherenkov-based LoC systems for compact in-line or at-line radionuclide monitoring downstream of appropriate sample preparation or radiochemical separation.
Abstract
Microfluidic lab-on-chip (LoC) systems miniaturise fluid handling and detection within compact devices, enabling at-line or in situ radiometric analysis while reducing sample handling and secondary waste. However, numerical optical modelling remains underexplored for guiding photon collection in radiometric microfluidic systems. Here, finite element ray tracing was used as a comparative design tool alongside experimental validation to develop a Cherenkov-based microfluidic platform for pure β-emitting radionuclides. Two Microfluidic Detection Units (MDUs), comprising single- and dual-channel serpentine geometries, were fabricated and integrated with a photomultiplier tube detection system. The model compared the effects of channel geometry, reflective backing, and detector coupling on rays reaching the detector. Experimental validation using 90Sr/90Y, 36Cl, and 99Tc assessed responses to multiple β-emitting radionuclides. Both MDUs demonstrated an approximately linear increase in count rate over the tested activity range (R2 > 0.96), while modelled and experimental responses showed consistent relative trends (R2 > 0.91). The single-channel MDU achieved the lower minimum detectable activity of 21 Bq, whereas the larger-volume dual-channel MDU achieved the lower activity-concentration threshold of 18.2 kBq L-1, compared with 26.3 kBq L-1 for the single-channel configuration. The second channel increased total detected signal, although not in proportion to the 3.5-fold increase in sample volume, indicating reduced photon collection from the deeper channel. This work demonstrates the value of modelling-guided optimisation for the development of radiometric microfluidic analytical platforms and highlights the potential of Cherenkov-based LoC systems for compact in-line or at-line radionuclide monitoring downstream of appropriate sample preparation or radiochemical separation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Microfluidic diffusional sizing in bioanalysis and biosensing.Lab on a chip · 2026
- Microfluidic Molecular Diagnostic Chips: Functional Modules, Integration Strategies, and Emerging Applications.Small methods · 2026
- Analytical performance of plastic scintillation microspheres versus a liquid scintillation cocktail for measuring beta-emitting radionuclides in representative aqueous media.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- A Portable UV-LED Dual-Nafion Microfluidic Platform for Label-Free Trace Protein Preconcentration and Absorbance Detection.Biosensors · 2026
- Early-stage nanoparticle-metal interactions under flow: real-time, sub-second monitoring by microfluidic-ICP-MS.Journal of colloid and interface science · 2026
- Lab-on-a-chip microfluidic devices for the analysis of trace metals: foundational principles, technological evolution, and translational prospects.Analytical methods : advancing methods and applications · 2026