Methodology for identification of depleted metallic uranium and natural metallic uranium by portable gamma spectrometer.
Samsul Arefin, S K Suman, N N Satvilkar, T I Khan, R V Kolekar
Health Physics Division, Bhabha Atomic Research Centre, Mumbai, 400085, India. · Health Physics Division, Bhabha Atomic Research Centre, Mumbai, 400085, India. Electronic address: sumansk@barc.gov.in.
Key points
From the paper's conclusion- The results demonstrate that, with proper calibration, and spectral analysis, portable gamma spectrometry can serve as an effective tool for real-time uranium isotope identification in applications ranging from nuclear safeguards and non-proliferation to environmental monitoring and emergency response.
Abstract
The requirement to distinguish between depleted metallic uranium (DU) and natural metallic uranium (NU) is essential in nuclear forensics, regulatory enforcement, nuclear fuel fabrication and environmental monitoring. Traditional laboratory-based methods offer high accuracy but are impractical for on-site measurement. This research develops a portable gamma spectrometry-based methodology for identifying DU and NU in real-time, non-destructively. The study utilizes a NaI(Tl) based gamma spectrometer to measure characteristic emissions from uranium isotopes and its daughter products. Key gamma peaks, particularly those at 92 keV (234Th) and 186 keV (235U), were analysed to derive isotopic ratios indicative of the material type. Emphasis was given on choosing right gamma peak and adjustment for x-ray contribution in 92 keV peak. Estimated activity ratio of 235U to 234Th of five natural uranium samples was greater than or equal to 0.046 and the activity ratio of five depleted uranium sample found to be less than 0.046, which validated the methodology. The results demonstrate that, with proper calibration, and spectral analysis, portable gamma spectrometry can serve as an effective tool for real-time uranium isotope identification in applications ranging from nuclear safeguards and non-proliferation to environmental monitoring and emergency response.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Integrated Prompt Fission Neutron and Gamma-Ray Logging for Direct Uranium Quantification in Exploration and In Situ Recovery MonitoringMinerals · 2026
- Development of a PGNAA facility based on compact D-T neutron generator for Er detection.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Development and application of nuclear forensic analytical capabilities at the Institute of Nuclear Chemistry and Technology.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Pilot full-scale demonstration of a prototype table-top neutron resonance transmission analysis system for nuclear material detectionCommunications Engineering · 2026
- Calibration of the NDUSRM series and further measurements of 236U/238U isotopic ratios in the Ewing collection.Journal of environmental radioactivity · 2026
- Analysis of an irradiated uranium sample for source attribution without chemical separation using microplasma ionization and ultrahigh resolution mass spectrometryJournal of Radioanalytical and Nuclear Chemistry · 2026