Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine

Development of a PGNAA facility based on compact D-T neutron generator for Er detection.

Wei Cheng, Can Cheng, Zhe Pan, Jiatong Li, Haoyu Lei, Pingkun Cai, Daqian Hei, Chenmin Xiao,

Department of Nuclear Science and Technology, Nanjing University of Aeronautics and Astronautics, Nanjing, 211106, China. · College of Electrical Engineering and Control Science, Nanjing Tech University, Nanjing, 211816, China.

Key points

From the paper's conclusion
  • This work confirms the feasibility of D-T neutron source-based PGNAA for Er detection and provides a validated framework for further optimization of facility design.

Abstract

A prompt gamma-ray neutron activation analysis (PGNAA) facility based on a compact D-T neutron generator and an HPGe detector was developed for the determination of burnable poisons in plate-type nuclear fuel elements. Erbium was selected as a representative neutron absorber due to its high thermal neutron capture cross section and its potential application in advanced fuel designs. Prompt gamma-ray spectra of an Er plate sample were measured, and the characteristic 0.816 MeV gamma line from the 167Er(n,γ)168Er reaction was clearly identified. The mass detection limit of Er was evaluated to be 31.07 g for a measurement time of 1000 s. A detailed Monte Carlo model based on GAMOS was established. The calculated prompt gamma-ray counts agreed with experimental results within 15%, supporting its use for preliminary response estimation. This work confirms the feasibility of D-T neutron source-based PGNAA for Er detection and provides a validated framework for further optimization of facility design.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers