An end-to-end quality assurance procedure for pelvic VMAT using optically stimulated luminescence dosimeters and an anthropomorphic phantom.
Ahlam Azalmad, Mohamed Abour, Mehdi El Ouartiti, Mohamed Hilal
Laboratory of Sciences and Engineering of Biomedicals Biophysics and Health, Higher Institute of Health Sciences, University Hassan Fisrt, Settat, Morocco. a.azalmad@uhp.ac.ma. · LSIB laboratory, Faculty of Sciences and Technologies Mohammedia, Hassan II University of Casablanca, Mohammedia, Morocco.
Key points
From the paper's conclusion- OSLD measurements in an anatomically realistic female pelvic phantom confirm that cervical cancer VMAT delivery is accurate and reproducible. These results support the reliability of Eclipse (v18) using the Acuros XB algorithm across high-dose, gradient, and low-dose regions.
Abstract
BACKGROUND: Volumetric modulated arc therapy (VMAT) enables highly conformal pelvic dose delivery, but its clinical benefit depends on agreement between the Treatment Planning System (TPS) calculation and the dose actually delivered in anatomically realistic conditions. PURPOSE: To experimentally validate VMAT dosimetric accuracy for cervical cancer using optically stimulated luminescence dosimeters (OSLDs) placed in an anthropomorphic female pelvic phantom. MATERIALS AND METHODS: A female pelvic phantom was CT-scanned. A dual-arc VMAT plan was created in Eclipse (v18) using the Acuros XB algorithm with heterogeneity correction. Fifty-three OSLDs were positioned across seven pelvic slices (28-34) spanning in-field, edge, peripheral, and out-of-field regions. Measured doses were compared to TPS values using a TG-119-style percent difference metric. RESULTS: Overall agreement was clinically acceptable. The mean %Diff (OSL vs. TPS) was + 3.56% (range - 2.13% to + 6.93%). In-field doses showed the closest agreement (≤ 3%). Larger deviations occurred at the field edge (≈ 3.17%-6.93%), consistent with steep dose gradients. Peripheral and out-of-field locations generally remained within ~ 5%-7%, where scatter and low-dose modeling uncertainty dominate. Reproducibility was strong: SD_OSL values were < 5.2%, with mean SD of 2.63% (in-field), 4.34% (edge), 4.44% (peripheral), and 4.96% (out-of-field). CONCLUSION: OSLD measurements in an anatomically realistic female pelvic phantom confirm that cervical cancer VMAT delivery is accurate and reproducible. These results support the reliability of Eclipse (v18) using the Acuros XB algorithm across high-dose, gradient, and low-dose regions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Independent Three-dimensional Dose Calculation from Log Files for Breast VMAT: A Measurement-free Quality Assurance Approach.Journal of medical physics · 2026
- Machine learning prediction of electronic-portal-imaging-device-based in vivo dosimetry results using plan complexity metrics, machine output and gamma passing rate: a feasibility study.Physical and engineering sciences in medicine · 2026
- Prediction of patient-specific quality assurance for volumetric modulated arc therapy using machine log files.Quantitative imaging in medicine and surgery · 2026
- Machine learning-based prediction of gamma passing rates in VMAT prostate external radiation therapy plans.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- End-to-end technical validation of multi-isocenter total body irradiation using an anthropomorphic phantom.Radiological physics and technology · 2026
- Predicting VMAT patient-specific quality assurance outcome using a machine learning workflow.Physical and engineering sciences in medicine · 2026