Comparative Evaluation of Power-Law and Viscoelastic Hemolysis Models in the FDA Nozzle Benchmark Geometry.
Alberto Girelli, Giulia Giantesio, Adélia Sequeira, Tomáš Bodnár
Dipartimento di Matematica e Fisica "N. Tartaglia", Università Cattolica del Sacro Cuore, Brescia, Italy. · Dipartimento di Matematica e Fisica "N. Tartaglia", Università Cattolica del Sacro Cuore, Brescia, Italy. giulia.giantesio@unicatt.it.
Key points
From the paper's conclusion- The effect of $$\tau _S$$ τ S differed between the two models: it reduced hemolysis in the power-law case but increased it in the viscoelastic formulation.
Abstract
PURPOSE: The mechanical damage of red blood cells (hemolysis) remains a key challenge in the design of blood-contacting medical devices. This study aims to compare the predictive performance of the empirical power-law model and the physiologically motivated viscoelastic model in the FDA nozzle benchmark geometry, with a focus on how model parameters and shear stress definitions influence hemolysis predictions. METHODS: Computational fluid dynamics (CFD) simulations were performed using a finite volume framework to resolve flow fields and shear stresses in the FDA nozzle. Both hemolysis models were implemented under identical geometry, mesh and boundary conditions. A parametric analysis was conducted by varying key model parameters, including the viscoelastic constants A, T 1 , T 2 , the shear stress threshold τ S , and the power-law parameters. Predictions were evaluated using both total shear stress and turbulent viscous shear stress (TVSS) formulations, and compared against experimental data. CONCLUSION: The viscoelastic model demonstrated strong sensitivity to the relaxation time T 2 and to the choice of stress definition. A quasi-linear dependence on parameter A was identified, enabling calibration with experimental results. In contrast, the power-law model was less sensitive to stress definitions but strongly dependent on parameter fitting. The effect of τ S differed between the two models: it reduced hemolysis in the power-law case but increased it in the viscoelastic formulation. BACKGROUND: The viscoelastic model offers a more mechanistic description of hemolysis but requires careful parameter calibration and is highly sensitive to stress definitions. The power-law model remains computationally efficient and easier to implement, though empirical by nature. Future research should refine parameter estimation and extend validation to more complex flow regimes, considering feedback effects of hemolysis on viscosity. Ultimately, CFD prediction of hemolysis remains a challenging unsolved problem.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Methodology for computational assessment of hemolysis in under-occlusive roller pumps operated at clinically relevant conditions.Bioengineering & translational medicine · 2026
- Influence of hematocrit-dependent blood viscosity on hemodynamics of mitral paravalvular leak.Acta of bioengineering and biomechanics · 2026
- Comparative study of turbulence models for hemolysis prediction and hydraulic performance in centrifugal blood pumps.Clinical biomechanics (Bristol, Avon) · 2026
- Design and characterization of a Taylor–Couette device for studying factors affecting turbulence-induced hemolysisPhysics of Fluids · 2026
- Geometry-Dependent Variability in Rotational Rheometry of Whole Blood: Implications for Flow Curve Measurement and Protocol Standardization.Methods and protocols · 2026
- Computational fluid dynamics-based modeling of drug dissolution in non-Newtonian media.International journal of pharmaceutics · 2026