Geometry-Dependent Variability in Rotational Rheometry of Whole Blood: Implications for Flow Curve Measurement and Protocol Standardization.
Nebojša Pavlović, Biljana Vučković, Dušan Sedlarević, Nebojša Kladar, Stana Tokić, Ana Stupin, Krešimir Šolić, Veljko Krstonošić
Department of Pharmacy, Faculty of Medicine, University of Novi Sad, 21000 Novi Sad, Serbia. · Department of Pathophysiology and Laboratory Medicine, Faculty of Medicine, University of Novi Sad, 21000 Novi Sad, Serbia.
Key points
From the paper's conclusion- This study evaluated geometry-dependent variability in whole blood rheometry by comparing a double-gap coaxial cylinder (CC27 DG) with two cone–plate geometries differing in size and cone angle (C50 1° and C35 2°).
Abstract
Hemorheology has clear physiological relevance and potential diagnostic value, emphasizing the need for reliable blood rheological measurements. Rotational rheometry is widely used for the characterization of blood flow behavior, yet the influence of measuring geometry on rheological parameters remains insufficiently investigated and standardized. This study evaluated geometry-dependent variability in whole blood rheometry by comparing a double-gap coaxial cylinder (CC27 DG) with two cone-plate geometries differing in size and cone angle (C50 1° and C35 2°). Blood samples from five healthy volunteers were analyzed using an identical shear protocol from 0 to 100 s-1, and flow curves were fitted using the Power law model. All geometries consistently demonstrated non-Newtonian, shear-thinning behavior. Significant overall geometry effects were observed for the consistency index K (p = 0.022) and Power law goodness-of-fit r (p = 0.015), whereas no significant overall effect was detected for n (p = 0.247). While no pairwise comparison reached statistical significance after Holm adjustment, CC27 DG showed the lowest mean K, the highest mean r, and the smoothest flow curves, and C50 1° produced smooth curves with a reduced sample volume requirement. These findings support explicit reporting of measuring geometry and its consideration in the standardization of hemorheological protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Mastering Rheology: A Strategic and Practical Guide for Empowering All UsersIndustrial & Engineering Chemistry Research · 2025
- Non-Newtonian blood rheology significantly alters hemodynamic predictions during cardiac looping: a computational study.Frontiers in bioengineering and biotechnology · 2026
- Comparative Evaluation of Power-Law and Viscoelastic Hemolysis Models in the FDA Nozzle Benchmark Geometry.Cardiovascular engineering and technology · 2026
- Hemodynamic and hemorheological variations in coronary arteries: impact of stenosis geometry and disease-driven viscosity changesInternational Journal of Numerical Methods for Heat & Fluid Flow · 2026
- Geometry-Dependent Wall Shear Stress in Standardized 50% Coronary Stenoses: Effects of Pressure and Hematocrit in a Parametric CFD StudyMathematical and Computational Applications · 2026
- Characterizing Red Blood Cell Rheological Parameters Using Automated Syllectometry: Conventional and Exploratory Multivariate Analyses.International journal of laboratory hematology · 2026