Design and characterization of a Taylor–Couette device for studying factors affecting turbulence-induced hemolysis
Moritz Wiegand, Aris Groh, Simon Bender, Körling Pölkow, Christian Navid Nayeri, Katharina Vellguth, Vera Lommel, Jan Brüning,
Key points
From the paper's conclusion- The Taylor–Couette-device is suitable for controlled study of turbulence-related hemolysis.
Abstract
Blood-perfused medical devices operate predominantly under turbulent flow conditions, associated with increased flow-induced hemolysis. However, current hemolysis models are based on laminar flow experiments and do not account for turbulence. This study establishes an experimental setup to quantify turbulence-induced hemolysis and enables the development of hypotheses regarding interactions between turbulent flow and erythrocytes. A Taylor–Couette-device with counter-rotating cylinders was developed, operating with 20 ml fluid volume and capable of generating, visualizing, and measuring laminar and turbulent flow properties at identical mean shear rates. Flow regimes were characterized using an image-based, custom-trained neural network to ensure consistent classification across fluids. Experiments were performed with water and water–dextran solution as mono-phase blood-analogue fluids at physiologic mean shear rates of 256, 396, and 537 s−1, with one laminar and three turbulent measurement points. Laser Doppler anemometry (LDA) provided flow characterization, while torque measurements at the inner cylinder were used to estimate energy dissipation. The system reliably achieved laminar and turbulent flow regimes. LDA measurements provided velocity profiles, turbulence intensities, and Reynolds shear stresses in good agreement with literature. The intended increase in turbulence intensity across the turbulent points was realized at 396 and 537 s−1 but not at 256 s−1. Torque-, wall shear stress-, and LDA-based estimates of dissipation and Kolmogorov length scales agree in order of magnitude, with a near-constant offset attributable to bearing friction and motor losses. The Taylor–Couette-device is suitable for controlled study of turbulence-related hemolysis. Higher mean shear rates will require a reduced gap width, and two-phase blood-analogue fluids should be considered in future studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Methodology for computational assessment of hemolysis in under-occlusive roller pumps operated at clinically relevant conditions.Bioengineering & translational medicine · 2026
- Comparative Evaluation of Power-Law and Viscoelastic Hemolysis Models in the FDA Nozzle Benchmark Geometry.Cardiovascular engineering and technology · 2026
- Comparative study of turbulence models for hemolysis prediction and hydraulic performance in centrifugal blood pumps.Clinical biomechanics (Bristol, Avon) · 2026
- Mixing with viscoelastic waves at low Reynolds numbers.Lab on a chip · 2026
- Computational fluid dynamics assessment of a triple-lumen hemodialysis catheter.The journal of vascular access · 2026
- A-277 Improving Sample Integrity: Preclinical Evidence Demonstrating Reduced Hemolysis with Novel Blood Collection DeviceClinical Chemistry · 2026