Comparison of breast-level surface dose between single-energy CT and fast kilovoltage-switching dual-energy CT under matched CTDIvol conditions: a chest phantom study.
Mako Yamada, Takanori Masuda, Norimi Nishiyama, Junichi Nakagawa, Nanami Usui, Hitoshi Takahata, Takashi Amano, Shinichi Arao,
Department of Radiological Technology, Tsuyama Chuo Hospital, Okayama, Japan. · Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare, 288 Matsushima, Kurashiki-City, Okayama, 701-0193, Japan. takanorimasuda@yahoo.co.jp.
Key points
From the paper's conclusion- This phantom study compared breast-level surface dose and image noise between single-energy CT (SECT) and fast kilovoltage-switching dual-energy CT (DECT) under matched volume CT dose index (CTDIvol) conditions.
Abstract
This phantom study compared breast-level surface dose and image noise between single-energy CT (SECT) and fast kilovoltage-switching dual-energy CT (DECT) under matched volume CT dose index (CTDIvol) conditions. A chest phantom was scanned using a 256-row CT scanner. Surface dose was measured with real-time dosimeters placed on the bilateral anterior chest wall at the breast level and corresponding bilateral posterior positions. Tube currents of 195 mA for SECT and 250 mA for DECT provided matched console-displayed CTDIvol values. Each protocol was repeated 20 times as technical replicates. Image noise was measured in five circular regions of interest at the bilateral anterior and posterior chest walls and heart. Mean surface dose was lower with DECT than SECT at the breast level (15.86 vs. 17.02 mGy; mean difference [DECT minus SECT], -1.16 mGy; 95% confidence interval [CI], -1.71 to -0.61; P < 0.001) and posterior surface (15.77 vs. 17.26 mGy; mean difference, -1.49 mGy; 95% CI, -2.21 to -0.77; P < 0.001). Mean image noise was 7.08 HU with DECT and 6.66 HU with SECT, with no statistically significant difference (mean difference, 0.42 HU; 95% CI, -0.17 to 1.01; P = 0.160). Under the tested conditions, DECT showed lower anterior and posterior surface doses than SECT performed without CT automatic exposure control. No statistically significant difference in image noise was observed, but this does not establish equivalent image quality. These findings should not be generalized to clinically optimized SECT protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The impact of breast shielding on radiation dose during computed tomography procedures.Radiation protection dosimetry · 2026
- Effect of 70-kVp chest CT on surface dose at the mammary gland level in a neonatal phantom under matched CTDIvol conditions.Radiological physics and technology · 2026
- Evaluation of a fast kV switching dual energy CT in deriving relative stopping power using tissue equivalent phantoms of various sizesMedical Physics · 2026
- Fast Patient-Specific Breast CT Dosimetry: 22-Fold Acceleration of Monte Carlo MGD EstimationarXiv · 2026
- Minimum recommended distances between cardiac implantable electronic devices and the applicator surface in low-energy intraoperative radiotherapy of the breast: an in vivo dosimetric study.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Evaluation of quantitative accuracy in 4D DCE‐bCT: Insights from a perfusion phantom studyMedical Physics · 2026