The impact of breast shielding on radiation dose during computed tomography procedures.
Salman Albeshan, Haila Nasser A Aloud
Radiological Sciences Department, College of Applied Medical Sciences, King Saud University, ASl Diriyah district, King Abdullah street, Riyadh, 11433, Saudi Arabia.
Key points
From the paper's conclusion- Lead shielding and ATCM effectively reduced surface breast radiation dose, particularly in regions dominated by scatter.
Abstract
To quantify and compare breast radiation dose across CT protocols using shielding and automatic tube current modulation (ATCM) in single- and dual-energy scanning modes. Surface breast dose was measured using a calibrated pencil ionization chamber placed in a female anthropomorphic phantom. Scans were performed on a 128-slice CT system across multiple anatomical regions using manual mA, ATCM, lead shielding, and single- and dual-energy modes. In single-energy abdominal CT, shielding reduced measured surface breast dose by 12.2%, and ATCM provided an additional 38.2% reduction. Dual-energy CT resulted in higher measured surface breast doses than single-energy, but shielding reduced dose by approximately 62%. The greatest reductions were observed in head and neck (63.7%) and facial bone scans (89.5%). Lead shielding and ATCM effectively reduced surface breast radiation dose, particularly in regions dominated by scatter. Dual-energy imaging increased dose, while combining shielding with ATCM provided the greatest reduction.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of breast-level surface dose between single-energy CT and fast kilovoltage-switching dual-energy CT under matched CTDIvol conditions: a chest phantom study.Radiological physics and technology · 2026
- Influence of collimation on radiation exposure during chest radiography.Radiation protection dosimetry · 2026
- Quantifying radiation exposure in contrast-enhanced stereotactic breast biopsy: considerations for dose optimizationEuropean Radiology Breast · 2026
- CT protocol optimisation for oncologic surveillance imaging: phantom-based comparison of multi-scan and single-scan acquisition strategies.Biomedical physics & engineering express · 2026
- Radiation Dose to the Breast in Occupationally Exposed Healthcare Professionals: A Case Series.Health physics · 2026
- Breast cancer staging and follow-up: Contributions of chest computed tomography and dosimetric analysis.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026