Minimum recommended distances between cardiac implantable electronic devices and the applicator surface in low-energy intraoperative radiotherapy of the breast: an in vivo dosimetric study.

Licia Toscano, Michele Avanzo, Stefano Fracon, Samuele Massarut, Loredana Barresi, Aurora Corso, Saleh El Ghassal, Lorenzo Vinante,

Medical Physics Department, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, Aviano, PN, Italy. Electronic address: licia.toscano@cro.it. · Medical Physics Department, Centro di Riferimento Oncologico di Aviano (CRO) IRCCS, Aviano, PN, Italy.

Preclinical · n=21

Key points

From the paper's conclusion
  • Intraoperative ultrasound combined with applicator-specific minimum recommended distances provides a practical tool to guide treatment decisions in patients with CIEDs undergoing low-energy breast IORT.

Abstract

PURPOSE: Clinical guidelines recommend keeping the dose to cardiac implantable electronic devices (CIEDs) below 2 Gy. In low-energy intraoperative radiotherapy (IORT) of the breast, the Intrabeam treatment planning system (TPS) estimates dose in water, without accounting for tissue attenuation, reduced scatter at the skin surface and backscatter from tungsten shielding. This study aimed to determine conservative minimum recommended distances between the CIED and the applicator surface based on in vivo dosimetry. MATERIALS AND METHODS: Thermoluminescent dosimeters (TLDs), cross-calibrated against an ionization chamber, were placed on the skin between the applicator and the infraclavicular region, close to the usual CIED location, in 21 patients undergoing breast IORT. Applicator-to-dosimeter distances were measured with intraoperative ultrasound, and dose-distance curves were fitted to identify the distance at which dose falls below 2 Gy (d<2Gy). RESULTS: Tungsten shielding increased the surface dose by approximately 22%, whereas without shielding the surface dose in water was on average 3% lower than predicted under full-scatter conditions. The median d<2Gy exceeded the TPS value by 3.1 mm (interquartile range -1.7 to 4.5 mm), with large inter-patient variability. Minimum recommended distances, defined as the upper 95% confidence limit of the regression of d<2Gy on applicator diameter, ranged from 29 mm (35 mm applicator) to 39 mm (50 mm applicator), compared with 22-24 mm calculated by the TPS. CONCLUSION: Intraoperative ultrasound combined with applicator-specific minimum recommended distances provides a practical tool to guide treatment decisions in patients with CIEDs undergoing low-energy breast IORT.

Read the paper