Effect of 70-kVp chest CT on surface dose at the mammary gland level in a neonatal phantom under matched CTDIvol conditions.
Nanami Usui, Takanori Masuda, Norimi Nishiyama, Junichi Nakagawa, Mako Yamada, Nagito Kuwada, Hitoshi Takahata, Takashi Amano,
Department of Radiological Technologist, Okayama Saiseikai General Hospital, 2-25, Kokutai-cho, Kita-ku, Okayama-shi, Okayama, 700-8511, Japan. · Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare, 288 Matsushima, Kurashiki-city, Okayama, 701-0193, Japan. takanorimasuda@yahoo.co.jp.
Key points
From the paper's conclusion- Under matched CTDIvol conditions, 70-kVp neonatal chest CT increased anterior mammary surface dose relative to 120-kVp CT in this phantom.
Abstract
To compare surface dose at the mammary gland level between 70- and 120-kVp non-contrast neonatal chest computed tomography (CT) under matched volume CT dose index (CTDIvol) conditions, a neonatal anthropomorphic phantom was scanned using a 256-row CT scanner. Real-time dosimeters were placed bilaterally on the anterior chest wall at the mammary gland level and on the posterior chest wall at the same axial level. For the 70-kVp protocol, the noise index was adjusted to match the console-displayed CTDIvol of the 120-kVp protocol. Each protocol was repeated 15 times, and image noise was measured on axial images. Anterior surface dose was significantly higher with 70 kVp than with 120 kVp (median [IQR], 5.67 [5.17-5.86] vs. 5.29 [4.84-5.43] mGy; p < 0.01). Posterior surface dose and image noise did not differ significantly. Under matched CTDIvol conditions, 70-kVp neonatal chest CT increased anterior mammary surface dose relative to 120-kVp CT in this phantom.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of breast-level surface dose between single-energy CT and fast kilovoltage-switching dual-energy CT under matched CTDIvol conditions: a chest phantom study.Radiological physics and technology · 2026
- Optimization of image quality and radiation dose in low-dose pediatric chest CT Using 80 kVp acquisition and variable ASiR-V reconstruction: A retrospective comparative study.Medicine · 2026
- The impact of arm position on image quality and radiation dose in pediatric toddler chest computed tomography: a phantom and clinical study.Frontiers in medicine · 2026
- Measured fetal-region dose in X-ray pelvimetry and low-dose computed tomography using a realistic pregnant phantom: a comparative study.Radiation protection dosimetry · 2026
- Ultra-low 70-kV cardiac CT angiography in neonates: a comparative dose-optimized approach with preserved image quality.Frontiers in cardiovascular medicine · 2026
- Comparing 70 kV and 120 kV photon-counting CT protocols for pediatric cardiac imaging in children weighing 10-30 kg: Radiation dose and image quality.Journal of cardiovascular computed tomography · 2026