Comparing 70 kV and 120 kV photon-counting CT protocols for pediatric cardiac imaging in children weighing 10-30 kg: Radiation dose and image quality.

Nolan H Dang, Wei Zhou, Gladys M Arguello Fletes, Jason P Weinman, LaDonna J Malone

University of Colorado School of Medicine, Aurora, CO, USA. Electronic address: nolan.dang@cuanschutz.edu. · Department of Radiology, University of Colorado School of Medicine, 13123 East 16th Avenue, B125, Aurora, CO, 80045, USA. Electronic address: wei.zhou@cuanschutz.edu.

Key points

From the paper's conclusion
  • In children weighing 10-30 kg, 70 kV ultra-high-pitch cardiac PCCT was associated with lower radiation dose than 120 kV, with no significant difference in CNR and median reader scores in the diagnostic range. A 120 kV protocol remains appropriate when metal-artifact reduction or advanced spectral reconstructions are needed.

Abstract

BACKGROUND: Scanning at 70 kV on photon-counting CT (PCCT) reduces radiation dose by 50% compared to 120 kV in infants under 10 kg without loss of diagnostic quality. Whether scanning at a lower kV is of benefit in larger children, in whom greater attenuation increases image noise, has not been determined. METHODS: We retrospectively compared ultra-high-pitch cardiac PCCT examinations in children weighing 10-30 kg acquired at 70 kV (n = 79) or 120 kV (n = 80), both at a CARE keV IQ level of 25. We measured signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in the ascending aorta and main pulmonary artery and compared radiation dose. Two pediatric cardiac radiologists rated image quality in five vascular structures on a 5-point Likert scale. RESULTS: Groups were similar in age and size. Neither CNR nor SNR differed significantly. Median Likert scores were at least 4 for four of five structures in both groups. Median CTDIvol was 0.49 vs 0.63 mGy (P < 0.001), and effective dose 0.65 vs 0.75 mSv (P < 0.001) for 70 vs 120 kV. CONCLUSION: In children weighing 10-30 kg, 70 kV ultra-high-pitch cardiac PCCT was associated with lower radiation dose than 120 kV, with no significant difference in CNR and median reader scores in the diagnostic range. A 120 kV protocol remains appropriate when metal-artifact reduction or advanced spectral reconstructions are needed.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers