Comparing 70 kV and 120 kV photon-counting CT protocols for pediatric cardiac imaging in children weighing 10-30 kg: Radiation dose and image quality.
Nolan H Dang, Wei Zhou, Gladys M Arguello Fletes, Jason P Weinman, LaDonna J Malone
University of Colorado School of Medicine, Aurora, CO, USA. Electronic address: nolan.dang@cuanschutz.edu. · Department of Radiology, University of Colorado School of Medicine, 13123 East 16th Avenue, B125, Aurora, CO, 80045, USA. Electronic address: wei.zhou@cuanschutz.edu.
Key points
From the paper's conclusion- In children weighing 10-30 kg, 70 kV ultra-high-pitch cardiac PCCT was associated with lower radiation dose than 120 kV, with no significant difference in CNR and median reader scores in the diagnostic range. A 120 kV protocol remains appropriate when metal-artifact reduction or advanced spectral reconstructions are needed.
Abstract
BACKGROUND: Scanning at 70 kV on photon-counting CT (PCCT) reduces radiation dose by 50% compared to 120 kV in infants under 10 kg without loss of diagnostic quality. Whether scanning at a lower kV is of benefit in larger children, in whom greater attenuation increases image noise, has not been determined. METHODS: We retrospectively compared ultra-high-pitch cardiac PCCT examinations in children weighing 10-30 kg acquired at 70 kV (n = 79) or 120 kV (n = 80), both at a CARE keV IQ level of 25. We measured signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in the ascending aorta and main pulmonary artery and compared radiation dose. Two pediatric cardiac radiologists rated image quality in five vascular structures on a 5-point Likert scale. RESULTS: Groups were similar in age and size. Neither CNR nor SNR differed significantly. Median Likert scores were at least 4 for four of five structures in both groups. Median CTDIvol was 0.49 vs 0.63 mGy (P < 0.001), and effective dose 0.65 vs 0.75 mSv (P < 0.001) for 70 vs 120 kV. CONCLUSION: In children weighing 10-30 kg, 70 kV ultra-high-pitch cardiac PCCT was associated with lower radiation dose than 120 kV, with no significant difference in CNR and median reader scores in the diagnostic range. A 120 kV protocol remains appropriate when metal-artifact reduction or advanced spectral reconstructions are needed.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pediatric Chest Photon-counting CT: Dose and Noise Behavior of Clinical Sn100 kVp and 90 kVp Protocols across Image Quality Levels - a Phantom Study.RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin · 2026
- Photon-Counting CT Enables Higher Image Quality at a Lower Radiation Dose Compared with a 256-Slice Energy-Integrating Detector CT in Pediatric Patients with Congenital Heart DiseaseDiagnostics · 2026
- Radiation Dose and Image Quality of Photon-Counting Detector CT for Multiphasic Pre-Transplant Abdominal Imaging in Young Children Aged ≤36 Months: A Propensity-Matched Study.The British journal of radiology · 2026
- Photon-counting CT in pediatric patients: A multi-institutional evaluation of protocols and radiation exposure in 4,772 patientsDiagnostic and Interventional Imaging · 2025
- Next-gen paediatric imaging: ultra-low dose chest photon-counting CT approaching chest radiograph doses.European radiology experimental · 2026
- Ultra-low 70-kV cardiac CT angiography in neonates: a comparative dose-optimized approach with preserved image quality.Frontiers in cardiovascular medicine · 2026