Ultra-low 70-kV cardiac CT angiography in neonates: a comparative dose-optimized approach with preserved image quality.

Xiaofang He, Chi Nie, Xiaoyan Zheng, Lei Zheng, Chenxiao Wang, Ruqi Fang, Yumin Zhong

Department of Radiology, Fujian Children's Hospital (Fujian Branch of Shanghai Children's Medical Center), College of Clinical Medicine for Obstetrics & Gynecology and Pediatrics, Fujian Medical University, Fuzhou, China. · Diagnostic Imaging Center, Shanghai Children's Medical Center, Shanghai Jiao Tong University School of Medicine, Shanghai, China.

Observational · n=125

Key points

From the paper's conclusion
  • Ultra-low tube voltage (70 kV) cardiac CT with weight-adjusted iodine delivery is feasible for neonatal imaging and can substantially reduce radiation exposure while preserving coronary artery image quality and visibility. This approach may help achieve a more balanced trade-off among image quality, radiation dose, and contrast administration in neonatal cardiac CT examinations.

Abstract

OBJECTIVE: To evaluate the feasibility of ultra-low tube voltage (70 kV) cardiac CT angiography (CTA) with weight-adapted iodine delivery in neonates and to compare image quality, coronary visibility, radiation exposure, and iodine delivery with those of a conventional 80-kV protocol. METHODS: This retrospective study included 125 neonates with congenital heart disease who underwent prospectively ECG-triggered cardiac CTA using a 256-row CT scanner. Patients were divided into an 80-kV group reconstructed with 50% adaptive statistical iterative reconstruction-V (ASIR-V) (n = 61) and a 70-kV group reconstructed with 70% ASIR-V (n = 64). Both protocols used standardized weight-adapted iodine delivery (450 mg I/kg). Subjective and objective image quality, coronary artery visibility, radiation dose, and iodine delivery were compared between groups. RESULTS: Overall interobserver agreement for subjective image quality assessment was excellent (κ = 0.815). Subjective image quality scores were comparable between the two groups across most coronary segments, with no significant differences in overall image quality (3.98 ± 1.04 vs. 3.95 ± 1.05, P = 0.545). Objective image quality parameters, including CT attenuation, image noise, signal-to-noise ratio, and contrast-to-noise ratio, also showed no significant differences between groups (all P > 0.05). Coronary artery visibility remained high in both groups, with overall visibility rates exceeding 89%. Importantly, distal coronary segments remained diagnostically assessable in approximately 75% of cases under the 70-kV protocol. Compared with the 80-kV protocol, the 70-kV protocol achieved a 24% reduction in effective radiation dose (0.32 ± 0.05 vs. 0.42 ± 0.06 mSv, P < 0.001) without increasing iodine delivery. CONCLUSIONS: Ultra-low tube voltage (70 kV) cardiac CT with weight-adjusted iodine delivery is feasible for neonatal imaging and can substantially reduce radiation exposure while preserving coronary artery image quality and visibility. This approach may help achieve a more balanced trade-off among image quality, radiation dose, and contrast administration in neonatal cardiac CT examinations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Ultra-low 70-kV cardiac CT angiography in neonates: a comparative dose-optimized approach with preserved image quality — Science Skim