Agreement of a Handheld Glucose Meter with Laboratory Serum Glucose Measurements in Dairy Cows
Dovilė Malašauskienė, Mindaugas Televičius, Gabija Lembovičiūtė, Greta Šertvytytė, Samanta Grigė, Akvilė Girdauskaitė, Melanie Christina Siebert, Vita Riškevičienė,
Key points
From the paper's conclusion- These findings show that the two measurement approaches are not interchangeable for individual-cow assessment.
Abstract
Blood glucose concentration is commonly used to assess energy metabolism in dairy cows. Handheld glucose meters allow rapid measurement under farm conditions, but their performance in cattle may differ from that obtained with laboratory methods. This study evaluated the agreement between capillary whole-blood glucose measured with the Accu-Chek® Instant handheld glucose meter and venous serum glucose measured with a laboratory biochemical analyzer. A total of 93 Holstein–Friesian animals provided 117 paired glucose measurements; 24 animals were sampled on two occasions. Capillary blood was collected from the vulvar mucosa and measured immediately, while venous blood was collected from the coccygeal vein and processed to serum for laboratory analysis. The mean glucose concentration was 3.27 ± 0.51 mmol/L in capillary whole blood and 3.44 ± 0.42 mmol/L in venous serum. The mean paired difference was 0.17 mmol/L (95% CI, 0.06–0.27 mmol/L; p = 0.0018). The correlation between methods was weak (r = 0.267), and Lin’s concordance correlation coefficient was 0.246. Bland–Altman analysis showed limits of agreement from −0.94 to 1.27 mmol/L. A significant proportional bias was also identified (β = −0.302; 95% CI, −0.562 to −0.042; p = 0.023). Categorical agreement was limited (Cohen’s κ = 0.241), and 47 of 117 paired measurements were classified into different glucose categories. These findings show that the two measurement approaches are not interchangeable for individual-cow assessment. The observed differences may reflect not only device performance but also differences in sampling site, sample matrix, and sample handling.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Whole-blood metabolic biomarkers improve detection of subclinical ketosis in postpartum grazing dairy cows.JDS communications · 2026
- Portable blood glucose meters do not reliably detect hypoglycemia in ferrets (Mustela putorius furo) compared with a point-of-care chemistry analyzer.American journal of veterinary research · 2026
- Assessing measurement uncertainty at laboratory network scale and its impact on diagnostic performance in the absence of a gold standard: application to ELISA tests for Coxiella burnetii in ruminantsbioRxiv · 2026
- Predicting classes of β-hydroxybutyrate concentrations in blood of dairy cows from milk mid-infrared spectra: Application of a weighted discriminant model addressing class imbalance.JDS communications · 2026
- Comparison of Blood Glucose Measurement Methods at Low Concentrations in Neonates at Risk for Hypoglycemia.The journal of applied laboratory medicine · 2026
- Construction and Evaluation of a Ketosis Prediction Model Based on Milk MIR Spectra and Feature Wavenumber Selection in Dairy Cows.Animals : an open access journal from MDPI · 2026