Near-infrared spectroscopy for estimating mixed venous oxygen saturation and cardiac output during off-pump coronary artery bypass grafting: a prospective observational study.
Chahyun Oh, Soomin Lee, Sujin Baek, Hyeyoung Choi, Man-Shik Shim, Sang-Jun Park, Yunseo Ku, Boohwi Hong,
Department of Anesthesiology and Pain Medicine, College of Medicine, Chungnam National University, Daejeon, Korea. · Department of Thoracic & Cardiovascular Surgery, Chungnam National University Hospital, Daejeon, Korea.
Key points
From the paper's conclusion- To evaluate whether jugular venous NIRS (rSjvO₂) and baseline-calibrated cerebral NIRS (cal-rScO₂) can serve as surrogates for mixed venous oxygen saturation (SvO₂), and whether these inputs can support continuous cardiac output (CO) estimation during off-pump coronary artery bypass grafting (OPCAB).
Abstract
To evaluate whether jugular venous NIRS (rSjvO₂) and baseline-calibrated cerebral NIRS (cal-rScO₂) can serve as surrogates for mixed venous oxygen saturation (SvO₂), and whether these inputs can support continuous cardiac output (CO) estimation during off-pump coronary artery bypass grafting (OPCAB). In this prospective observational study, 25 adults undergoing elective OPCAB were enrolled at a single tertiary center. NIRS sensors were placed over the left internal jugular vein (IJV) under ultrasound guidance and over the bilateral forehead. Pulmonary artery catheter measurements served as the reference for SvO₂ and thermodilution CO (COTD). Agreement was assessed using Bland-Altman analysis with linear mixed-effects modeling, and trending ability by error grid analysis. CO was estimated using a previously derived modified Fick equation. rSjvO₂ showed modest bias versus SvO₂ (2.0%) but wide limits of agreement (LoA, - 14.9 to 18.9%). Accuracy varied with jugular anatomy, with the lowest error at an IJV depth of 1.38 cm. Cal-rScO₂ showed better agreement (bias 0.8%; LoA, - 7.8 to 9.4%), excellent discrimination for SvO₂ <65% (AUC, 0.970), and good trending ability. In contrast, NIRS-based CO estimation showed significant proportional bias, 33.0% percentage error (95% CI, 27.4 to 41.5%), and poor trending ability. Cal-rScO₂ appears to be a useful surrogate for SvO₂ trending during OPCAB after simple one-point calibration. Jugular NIRS may be useful in selected patients, whereas NIRS-based CO estimation remained clinically insufficient.Trial registration: Clinical Research Information Service (CRIS), Republic of Korea (KCT0007918 November 18, 2022).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The effect of near-infrared spectroscopy guided intraoperative care on neurocognitive and functional outcomes after cardiac surgery: a systematic review and meta-analysis of randomised controlled trials.British journal of anaesthesia · 2026
- Intraoperative monitoring of myocardial oxygenation during open-heart surgery: feasibility and initial performance.Journal of clinical monitoring and computing · 2026
- Inline dissolved-gas-derived oxygen saturation reference for dynamic blood phantoms at adult-head optical properties.Biomedical optics express · 2026
- Effects of patient demographic and clinical data on the assessment of cerebral oxygen monitoring values during coronary artery bypass grafting.Perfusion · 2026
- Microvascular oxygenation responses are associated with the reactive hyperemic stimulus but not brachial artery flow-mediated dilation.Microvascular research · 2026
- Near-Infrared Spectroscopy-Guided Selective Distal Perfusion in Femoral Venoarterial Extracorporeal Membrane Oxygenation: A Feasibility Study.Diagnostics (Basel, Switzerland) · 2026