Evaluation of a rapid batch export system for electroretinography waveform data.
Aomeng Cui, Maria I Nunez, Mia OConnell, Karen M Chen, Stanley Chang, Leejee Suh, Scott E Brodie, Vlad Diaconita
Vagelos College of Physicians and Surgeons, Columbia University Irving Medical Center, 622 W 168th Street, Presbyterian Hospital Building, 18thFloor, New York, NY, 10032, USA. · Vagelos College of Physicians and Surgeons, Columbia University Irving Medical Center, 622 W 168th Street, Presbyterian Hospital Building, 18thFloor, New York, NY, 10032, USA. vd2374@cumc.columbia.edu.
Key points
From the paper's conclusion- Use of batch export with manual oversight for sources of potential error can allow for faster, more effective data gathering and analysis of large-scale ERG waveform datasets in academic and clinical trial settings.
Abstract
PURPOSE: Accurate application of electroretinography (ERG) remains challenging due to the heterogeneity of retinal disorders. Despite its potential as a diagnostic tool, the clinical use of ERG is limited by the expertise required for interpretation and the error-prone manual extraction of waveform data. At scale, waveform analysis in academic research and clinical trials is made difficult by these challenges. This study performed external validation and quality assessment of automated compared to manual extraction and collection of raw ERG waveforms. METHODS: Raw waveform data was collected using International Society for Clinical Electrophysiology of Vision (ISCEV) standard ERG studies from the ophthalmology practice at Columbia University. A recently developed automated batch exporting tool (Espion, Diagnosys LLC, Lowell, MA) was used to retrieve and compile data and compare with manually collected data for 357 patient sessions. RESULTS: The automated and manually exported data had 99.75% ± 0.17% similarity using a Jaccard similarity test, suggesting the automated tool is both rapid and highly accurate. We identified several inconsistencies leading to errors in the automated batch export. The most common errors in compiled batch exported data included: improperly merging two ERG files taken on the same day and lack of accounting for missing data. CONCLUSIONS: Use of batch export with manual oversight for sources of potential error can allow for faster, more effective data gathering and analysis of large-scale ERG waveform datasets in academic and clinical trial settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- General-purpose time-series foundation models enable sample-efficient transfer learning in retinal electrophysiologymedRxiv · 2026
- Bit-Exact Reconstruction and Certification of Electrocardiogram Waveforms from Vector-Encoded PDF FilesmedRxiv · 2026
- Comparison of the Melbourne rapid fields-web device-independent online perimeter and the octopus 600 perimeter in glaucoma patients.Frontiers in ophthalmology · 2026
- Comparison of raw-signal, feature-based and hybrid representations for machine-learning classification of retinal diseases from pattern electroretinography.Medical & biological engineering & computing · 2026
- Feasibility and performance of eye movement perimetry for glaucoma triage in Tanzania: A prospective cross-sectional study in a real-world setting.PLOS global public health · 2026
- Analytical Agreement Between the Vision C Automated ESR Analyzer and the Manual Westergren Reference Method: A Prospective Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026