Hemodynamic Analysis of a Repaired Ascending Aorta with Preserved Aortic Root.
Hannah Zhai, Yurui Chen, Yuichiro Kitada, Hiroo Takayama, Vijay Vedula
Department of Mechanical Engineering, Columbia University, New York, NY, 10027, USA. · Division of Cardiothoracic and Vascular Surgery, Department of Surgery, New York Presbyterian Hospital/Columbia University Irving Medical Center, New York, NY, 10032, USA.
Key points
From the paper's conclusion- We demonstrated that restoring a normal sinus shape and STJ profile during ascending aortic repair may have potential hemodynamic benefits. The study also highlights the model's sensitivities, particularly the LVOT shape and leaflet morphology and mobility, and may assist planning STJ reconstruction to yield optimal hemodynamics before intervention.
Abstract
PURPOSE: To evaluate the hemodynamic impact of restoring a normal sino-tubular junction (STJ) following a novel Hegar dilator-based procedure in patients undergoing root-sparing ascending thoracic aortic aneurysm (ATAA) repair using computational modeling. METHODS: We retrospectively selected an ATAA patient who underwent pre- and postoperative gated computed tomography angiography (CTA). We developed a novel workflow to segment the lumen, thick-walled aorta, and aortic valve from CTA images for subsequent blood flow analysis using computational fluid dynamics (CFD) and fluid-structure interaction (FSI). Morphological and hemodynamic characteristics of the root were quantified and compared against those of a control subject. Sensitivity analyses were performed to evaluate the impact of graft properties, leaflet material parameters and heterogeneity, wall compliance, external tissue support, and solid domain annular boundary conditions. RESULTS: Both CFD and FSI results showed that the postoperative geometry reconstructed with a normal STJ profile reintroduces sinus vortices during peak systole, similar to the control subject, which were absent pre-surgery. Accounting for aortic valve leaflets in FSI studies yielded a similar spatial patterning of the high-velocity jet and wall shear stress compared to the CFD data. However, FSI resulted in substantially elevated velocities, time-averaged wall shear stress (TAWSS), and energy dissipation, likely due to the dynamically changing orifice area and differing profiles of the left ventricular outflow tract (LVOT). CONCLUSION: We demonstrated that restoring a normal sinus shape and STJ profile during ascending aortic repair may have potential hemodynamic benefits. The study also highlights the model's sensitivities, particularly the LVOT shape and leaflet morphology and mobility, and may assist planning STJ reconstruction to yield optimal hemodynamics before intervention.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- From Case Studies to Cohort of Patients: Automating FSI Simulations to Uncover Downstream Effects of Ascending Aortic Grafts.Annals of biomedical engineering · 2026
- Realistic anatomical model reconstruction and high-fidelity flow computation in stented coronary arteries for the analysis of in-stent thrombosis.Frontiers in cardiovascular medicine · 2026
- Fully coupled patient-specific fluid-structure interaction modeling of post-TAVI hemodynamics compared with echocardiographic measurements.Biomechanics and modeling in mechanobiology · 2026
- Impact of Bridging Stent Flaring Angle on Hemodynamic Performance-In Vitro Analysis.Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2026
- Fluid-structure interaction analysis of aortic valve dynamics: effects of aspect ratio.Regenerative biomaterials · 2026
- Patient-specific hemodynamic assessment using experimental and computational approaches in anomalous aortic origin of the coronary artery: Case study.JTCVS techniques · 2026