Impact of Bridging Stent Flaring Angle on Hemodynamic Performance-In Vitro Analysis.

Natasha Hasemaki, Jan Stana, Daniel Becker, David Khangholi, Nikolaos Konstantinou, Nikolaos Tsilimparis

Department of Vascular Surgery, University Hospital LMU Munich, Germany.

Preclinical

Key points

From the paper's conclusion
  • Within this controlled renal model, greater BSG protrusion length was associated with more pronounced local hemodynamic disturbances, whereas no consistent differences were observed among the flaring angles tested at a 3-mm protrusion. These hypothesis-generating findings support limiting unnecessary intraluminal protrusion and warrant validation in anatomically diverse models using clinically available BSGs.Clinical ImpactThis study provides a mechanistic basis for optimizing bridging stent-graft (BSG) deployment during fenestrated endovascular aortic repair. In this controlled in vitro model with four-dimensional flow magnetic resonance imaging and vector flow ultrasound, protrusion lengths ≥6 mm generated increased wall shear stress, oscillatory shear index, and sustained flow recirculation, whereas variations in flaring angle had minimal impact. These findings suggest that minimizing unnecessary aortic protrusion during implantation may reduce hemodynamic stress at the stent-aorta interface. Because idealized surrogates were used, the results are hypothesis-generating and do not support conclusions regarding the performance of individual commercially available BSGs.

Abstract

OBJECTIVE: In fenestrated endovascular aortic repair, bridging stent-grafts (BSGs) ensure target vessel (TV) incorporation, yet their geometric configuration may influence local hemodynamics and TV durability. This experimental study evaluated the relative impact of BSG protrusion length and flaring angle on aortic hemodynamics. METHODS: An idealized 3D-printed abdominal aortic model with bilateral renal arteries was integrated into an MRI-compatible mock circulatory loop reproducing physiological pulsatile flow. Seven idealized, non-commercial BSG configurations were investigated. Flaring angle was varied (35°, 60°, 70°, 85°) at a fixed 3-mm protrusion, and protrusion length (4, 6, 10 mm) was subsequently varied at a fixed flaring angle of 85°. Hemodynamic assessment was performed using 4-dimensional flow magnetic resonance imaging and vector flow ultrasound. Qualitative flow patterns, wall shear stress (WSS), and oscillatory shear index (OSI) were analyzed proximal and distal to the aortic outlet of the BSG. RESULTS: Qualitative analysis demonstrated localized flow separation and recirculation zones at the aortic outlet of BSGs, increasing with greater protrusion length. Protrusion lengths of 3 and 4 mm showed limited flow disturbance, whereas protrusion lengths of 6 and 10 mm were associated with pronounced vortex formation and sustained recirculation. Quantitative analysis revealed increased WSS and OSI with longer protrusion lengths, independent of flaring angle. Variations in flaring angle did not result in consistent changes in flow topology or shear metrics. Flow disturbances resolved downstream within 20-30 mm, depending on protrusion length. CONCLUSION: Within this controlled renal model, greater BSG protrusion length was associated with more pronounced local hemodynamic disturbances, whereas no consistent differences were observed among the flaring angles tested at a 3-mm protrusion. These hypothesis-generating findings support limiting unnecessary intraluminal protrusion and warrant validation in anatomically diverse models using clinically available BSGs.Clinical ImpactThis study provides a mechanistic basis for optimizing bridging stent-graft (BSG) deployment during fenestrated endovascular aortic repair. In this controlled in vitro model with four-dimensional flow magnetic resonance imaging and vector flow ultrasound, protrusion lengths ≥6 mm generated increased wall shear stress, oscillatory shear index, and sustained flow recirculation, whereas variations in flaring angle had minimal impact. These findings suggest that minimizing unnecessary aortic protrusion during implantation may reduce hemodynamic stress at the stent-aorta interface. Because idealized surrogates were used, the results are hypothesis-generating and do not support conclusions regarding the performance of individual commercially available BSGs.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers