Fully coupled patient-specific fluid-structure interaction modeling of post-TAVI hemodynamics compared with echocardiographic measurements.
Chiara Catalano, Deniz Ozturk, Maria Bastron, Alessandra Zerillo, Stefano Cannata, Caterina Gandolfo, Salvatore Pasta
Department of Research, IRCCS ISMETT, Via Tricomi, 5, Palermo, Italy. · Medividia, Herent, Belgium.
Key points
From the paper's conclusion- The proposed fully coupled FSI framework enables patient-specific analysis of post-TAVI hemodynamics and valve mechanics for S3 balloon-expandable intra-annular system.
Abstract
To develop a fully coupled, patient-specific fluid-structure interaction (FSI) framework for quantitative assessment of post-transcatheter aortic valve implantation (TAVI) hemodynamics and valve biomechanics, and to compare selected simulation-derived hemodynamic indices with post-procedural echocardiographic measurements. Patient-specific geometries were reconstructed from pre-operative computed tomography angiography in five subjects treated with SAPIEN 3 Ultra (S3) devices. Structural TAVI deployment was simulated using Abaqus/Explicit and subsequently coupled with FlowVision for performing a two-way post-TAVI FSI analysis. Personalized boundary conditions were derived from clinical measurements, including heart rate, blood pressure, and echocardiographic flow data. Predicted peak velocity, effective orifice area (EOA), and transvalvular pressure gradients (TPG) were quantitatively compared with post-procedural echocardiography using empirical cumulative distribution functions and area-based error metrics. The FSI framework reproduced realistic leaflet kinematics and patient-specific flow patterns, highlighting marked inter-patient variability despite identical device types. Average of predicted peak systolic velocity (2.66 ± 0.59 m/s) and TPG (22 ± 10.7 mmHg) showed good agreement with echocardiographic measurements as the area metric was below 10% for both peak TPG and velocity. Larger discrepancies were observed for EOA due to patient variability and reliability of echocardiographic measurements. The model additionally quantified biomechanical and hemodynamic parameters, including leaflet stress, time-averaged wall shear stress (TAWSS), and blood residence time (BRT). The proposed fully coupled FSI framework enables patient-specific analysis of post-TAVI hemodynamics and valve mechanics for S3 balloon-expandable intra-annular system. The preliminary comparison with echocardiographic measurements showed good agreement for peak velocity and peak TPG, while larger EOA discrepancies highlight the need for further clinical benchmarking using larger cohorts and more direct measurement modalities.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimental Bench Validation of a Computational Fluid-Structure Interaction Model of Transcatheter Aortic Valve DynamicsarXiv · 2026
- In Silico Modeling and Validation of Post-Dilation in TAVI Patients.Annals of biomedical engineering · 2026
- In silico modeling of transcatheter heart valve oversizing and ellipticity, Part I: Establishing credibility of an advanced modelComputer Methods and Programs in Biomedicine · 2026
- An extended fluid-structure interaction framework for aortic hemodynamics and cardiovascular risk assessment using physics-informed neural networks.Biomechanics and modeling in mechanobiology · 2026
- Construct validity of a fibre-optic pressure sensor integrated TAVI guidewire for invasive pressure measurements in transcatheter aortic valve implantation: a prospective registryEuropean Heart Journal · 2026
- In Vitro Hemodynamic Assessment of a Novel Pulmonary Visceral Pleura (PVP) Transcatheter Aortic Valve Replacement Prototype Based on Optical and Acoustic Measurements.Bioengineering (Basel, Switzerland) · 2026