Impact of urine concentration and automatic interpretation on sensitivity and specificity of Legionella BinaxNOW® results.
Camille Jacqueline, Ghislaine Descours, Anne-Gaëlle Ranc, Coralie Espinasse, Sophie Jarraud, Laetitia Béraud
Centre National de Référence des Légionelles, Hospices Civils de Lyon, Institut des Agents Infectieux, Lyon, France. · Hospices Civils de Lyon, Lyon, France.
Key points
From the paper's conclusion- Both urine concentration and automated interpretation with the Digival™ reader increased the sensitivity of the BinaxNOW® assay. Automated interpretation achieved this without the additional processing step required for concentration and may therefore represent a practical alternative. Heat treatment of positive samples remains necessary to maintain optimal specificity.
Abstract
PURPOSE: Urinary antigen tests (UATs) are widely used for rapid diagnosis of Legionnaires' disease (LD), primarily detecting Legionella pneumophila serogroup 1 lipopolysaccharide (LPS). Strategies, such as urine concentration, were shown to improve performance but mostly in retrospective studies. METHODS: We evaluated the impact of urine concentration on the diagnostic sensitivity and specificity of the BinaxNOW® assay. A set of 90 positive and 200 negative urine samples were prospectively collected. Another set of 48 positive urine samples were selected retrospectively to evaluate both urine concentration but also automated interpretation using the Digival™ reader on test sensitivity and specificity. RESULTS: We found that urine concentration significantly increased sensitivity by 9% before heat treatment. Automated interpretation increased sensitivity by 16.3% in unconcentrated samples. Before heat treatment, specificity remained high with visual interpretation (100% unconcentrated) but slightly decreased with automated reading (98% unconcentrated). After heat treatment, specificity reached 100% for all methods. In LPS-spiked samples, automated interpretation improved analytical sensitivity by 2- to 10-fold, particularly for mAb 3/1-positive strains. CONCLUSION: Both urine concentration and automated interpretation with the Digival™ reader increased the sensitivity of the BinaxNOW® assay. Automated interpretation achieved this without the additional processing step required for concentration and may therefore represent a practical alternative. Heat treatment of positive samples remains necessary to maintain optimal specificity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 62 Optimization of Legionella pneumophila Diagnostic Testing in a Large Healthcare System in Southeast TexasAmerican Journal of Clinical Pathology · 2026
- Analytical validation of a highly accurate and reliable next-generation sequencing-based urine assayMicrobiology Spectrum · 2026
- Can Digital PCR Enhance the Detection of Legionella pneumophila in Legionnaires' Disease Cases?Pathogens (Basel, Switzerland) · 2026
- B-001 Limitations of Urine Dipstick Testing and the Clinical Value of Creatinine-Normalized Quantitative Urinalysis in Urologic and Renal Disease DetectionClinical Chemistry · 2026
- B-303 Evaluating an Automated Urinalysis System: Analytical Performance and Verification StrategiesClinical Chemistry · 2026
- Rapid direct-from-urine growth detection to triage culture-negative urine: a prospective diagnostic accuracy study of a two-tube nitrate-reduction and bromothymol blue assay (SIBURI/SIBURI+)medRxiv · 2026