Clinical outcomes of a new presbyopia-correcting intraocular lens with spiral optic.
Alfonso Arias Puente, Amparo Carrero Collantes, José Luis Rincón Rosales, Lucía Infiesta-Valderrey, David P Piñero
Department of Ophthalmology, University Hospital Vithas Arturo Soria, Madrid, Spain. · Department of Ophthalmology, Vithas Medimar International Hospital, Alicante, Spain.
Key points
From the paper's conclusion- The spiral optic-based IOL provides a continuous full range of vision with optimized CS and a controlled incidence of visual disturbances, leading to high levels of patient satisfaction.
Abstract
PURPOSE: To analyse the visual performance, optical quality, and patient satisfaction following bilateral implantation of a new intraocular lens (IOL) with an spiral optic design. METHODS: Prospective, observational, non-comparative study including a total of 98 eyes of 49 patients (age range: 43-79 years). All eyes underwent uneventful bilateral cataract surgery with the implantation of the Galaxy IOL (Rayner®, Worthing, UK). Visual acuity (VA), refraction, mesopic and photopic contrast sensitivity (CS), defocus curve, simulated halos and glare perception, and patient-reported outcomes (Catquest 9SF questionnaire) were evaluated during a 3-month follow-up. RESULTS: Mean 3-month postoperative spherical equivalent and cylinder were - 0.06 ± 0.26 and - 0.12 ± 0.25 D. Mean binocular 3-monh postoperative logMAR corrected distance VA was - 0.08 ± 0.06, and mean 3-month binocular distance-corrected intermediate (66 cm) and near (40 cm) VAs were - 0.06 ± 0.06 and 0.11 ± 0.08 logMAR, respectively. The binocular defocus curve at 1 and 3 months showed functional vision (≤ 0.2 logMAR) from + 1.00 D to -2.50 D. Mesopic and photopic CS showed consistent improvement between 1 and 3 months almost across all spatial frequencies (p ≤ 0.035). Likewise, a marked reduction in halo and glare phenomena was observed from 1 to 3 months postoperatively (p < 0.001). Rasch Catquest scores showed a significant reduction in self-reported visual disability at both 1 month and 3 months postoperatively compared to baseline (p < 0.001). CONCLUSIONS: The spiral optic-based IOL provides a continuous full range of vision with optimized CS and a controlled incidence of visual disturbances, leading to high levels of patient satisfaction.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance of the First Spiral Refractive Intraocular Lens for Continuous Full Range of VisionJournal of Refractive Surgery · 2025
- 12-Month Visual and Clinical Outcomes After Implantation of a Monofocal Aspheric Hydrophobic Acrylic IOL: A Prospective Feasibility Study.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Visual Performance of a Novel Intraocular Lens With Higher-Order Aspheric Design Optimized for Intermediate Vision.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Impact of pupil size on defocus curve, contrast sensitivity and patient satisfaction in a diffractive, simultaneous vision intraocular lensJournal of Cataract & Refractive Surgery · 2026
- Straylight-Based Evaluation of Glare in Two Presbyopia-Correcting Intraocular Lenses.Klinische Monatsblatter fur Augenheilkunde · 2026
- Clinical Outcomes of a Novel Customized Multifocal Corneo-Scleral Contact Lens With Spherical Aberration Control.Eye & contact lens · 2026