Repeatability and reproducibility of corneal endothelial measurements using the topcon SP-1P specular microscope.
Rubén D Berrospi, Carolina M Estrella, Virgilio Galvis, Sylvia J Villamizar, Juan F Urrea, Alejandro Tello
Fundación Oftalmológica de Santander (FOSCAL), Calle 157 # 20 - 94, Torre C, Consultorio 301 Cañaveral, Floridablanca, Santander, Colombia. · Fundación Oftalmológica de Santander (FOSCAL), Calle 157 # 20 - 94, Torre C, Consultorio 301 Cañaveral, Floridablanca, Santander, Colombia. alejandrotello@gmail.com.
Key points
From the paper's conclusion- The Topcon SP-1P demonstrates high reproducibility for primary endothelial parameters, particularly when using the panoramic acquisition mode. Since systematic differences exist between modalities, single and panoramic measurements are not directly interchangeable, highlighting the importance of standardizing techniques for longitudinal clinical monitoring.
Abstract
PURPOSE: To evaluate the repeatability and reproducibility of corneal endothelial parameters measured with the Topcon SP-1P specular microscope in healthy eyes, comparing single and panoramic acquisition modes. METHODS: This reproducibility study included 137 eyes from 79 healthy adults. Endothelial cell density (ECD), average cell area, coefficient of variation (CV), hexagonality, and pachymetry were measured twice by two independent examiners using both single and panoramic modes. Statistical agreement was assessed using intraclass correlation coefficients (ICC), repeatability coefficients (RC), and Bland-Altman analysis. RESULTS: Intra-observer reproducibility for ECD, average cell area, and pachymetry was excellent in both modes (ICC ≥ 0.95). However, the panoramic mode consistently showed improved precision, with higher ICCs and lower RCs for most morphometric variables. Inter-observer agreement was also excellent for ECD (ICC 0.94) and average cell area (ICC 0.95). Significant systematic differences were found between acquisition modalities; specifically, ECD was on average 7.27% lower in single mode compared to panoramic mode (p < 0.001). CONCLUSION: The Topcon SP-1P demonstrates high reproducibility for primary endothelial parameters, particularly when using the panoramic acquisition mode. Since systematic differences exist between modalities, single and panoramic measurements are not directly interchangeable, highlighting the importance of standardizing techniques for longitudinal clinical monitoring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Establishing a precision reference benchmark for corneal epithelial aberration measurements using a new hybrid device across keratoconus stages.Eye and vision (London, England) · 2026
- Agreement of corneal thickness measured by Scheimpflug, SS-OCT, and specular microscopy in Chinese myopia.BMC ophthalmology · 2026
- Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy EyesPhotonics · 2026
- The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Determinants of posterior corneal optical densitometry and their relationship with endothelial morphology.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026
- Central corneal endothelial morphometry in emmetropic adults aged 20-40 years examined in Valencia, Spain: a retrospective cross-sectional analysis.Surgical and radiologic anatomy : SRA · 2026