European radiology

Angio-CT does not increase patient radiation exposure compared to conventional cone-beam CT for intra-arterial liver procedures.

Charlotte Schneider, Han Keijzers, Erik Jan Rijkhorst, Kenneth Hergaarden, Tarik Baetens, Regina Beets-Tan, Belinda de Wit-van der Veen, Elisabeth Klompenhouwer

Department of Radiology, The Netherlands Cancer Institute, Amsterdam, The Netherlands. · Department of Medical Physics and Technology, The Netherlands Cancer Institute, Amsterdam, The Netherlands.

Observational

Key points

From the paper's conclusion
  • Hybrid angiography suites incorporating angio-CT significantly reduced 3D effective dose compared to single-phase CBCT imaging. However, the total effective dose did not remain significant after uncertainty correction. Therefore, the use of angiography suites incorporating dual-phase angio-CT does not appear to change patient radiation exposure during intra-arterial liver procedures.

Abstract

OBJECTIVES: To evaluate patient radiation dose during intra-arterial liver procedures using Computed Tomography angiography (angio-CT) compared with C-arm Cone-Beam CT (CBCT). MATERIAL AND METHODS: This retrospective study included bland embolization and radioembolization mapping procedures performed between November 2021 and December 2024 using dual-phase angio-CT or single-phase CBCT on a hybrid angiography suite. Effective dose was calculated by converting dose-area and dose-length product using conversion factors. Fluoroscopy-only (2D), CBCT/angio-CT (3D) effective dose and total effective dose were compared using the Mann-Whitney U-test. Bootstrap resampling and permutation testing with Monte Carlo error propagation were applied to assess the robustness of 3D and total effective dose comparisons. RESULTS: Ninety-two procedures were included: 48 in the CBCT group and 44 in the angio-CT group. The 2D effective dose was comparable. The 3D effective dose was 4.1-fold lower in the angio-CT group (median 9.85 [IQR: 5.37-22.6 mSv] vs 39.9 mSv [IQR: 21.2-69.6 mSv], p < 0.001), resulting in 1.6-fold reduction in total effective dose for procedures performed by angio-CT (median 34.6 [IQR: 23.7-54.3 mSv] vs 55.9 mSv [IQR: 37.6-103 mSv], p < 0.001). After dose conversion uncertainty correction, corrected p-values were < 0.001 for 3D effective dose and 0.053 for the total dose. CONCLUSION: Hybrid angiography suites incorporating angio-CT significantly reduced 3D effective dose compared to single-phase CBCT imaging. However, the total effective dose did not remain significant after uncertainty correction. Therefore, the use of angiography suites incorporating dual-phase angio-CT does not appear to change patient radiation exposure during intra-arterial liver procedures. KEY POINTS: Question What is the effect of integrating angio-CT into hybrid angiography suites on patient radiation exposure? Findings Dual-phase angio-CT integration into intra-arterial liver therapy workflows does not increase overall procedural radiation exposure and is associated with a decreased 3D effective dose. Clinical relevance Angio-CT offers several clinical advantages, and existing literature demonstrates improved image quality compared with CBCT. This study shows that dual-phase angio-CT does not increase patient radiation exposure compared to single-phase CBCT, supporting its use when available.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Angio-CT does not increase patient radiation exposure compared to conventional cone-beam CT for intra-arterial liver procedures — Science Skim