Analytical and bioanalytical chemistry

Evaluation and correction of charcoal-stripped serum matrix bias in stable-isotope-dilution LC-MS/MS quantification of endogenous serum folates.

Shuangshuang Zhu, Yik Lam Cho, Ying Deng, Meng Wei, Chan Weng Huen, Chui Mae Wong, Joo Guan Yeo, Xueming Dong

School of Biological Science, Nanyang Technological University, Singapore, 637551, Singapore. · Translational Immunology Institute, SingHealth Duke-NUS Academic Medical Centre, Singapore, 169856, Singapore.

Key points

From the paper's conclusion
  • These findings show that CSS-based stable-isotope-dilution LC-MS/MS can produce residual matrix-dependent bias despite normalization using stable-isotope-labeled internal standards (SIL-IS), and that pooled-serum standard addition provides a practical correction strategy.

Abstract

Isotope-dilution LC-MS/MS is widely used for targeted quantification of endogenous metabolites, but surrogate-matrix calibration can remain biased when the surrogate matrix is noncommutable with native specimens. In this study, three endogenous serum folate vitamers were quantified by stable-isotope-dilution LC-MS/MS using charcoal-stripped serum (CSS) as a surrogate blank matrix. Post-extraction spiking and pooled-serum standard addition showed analyte-dependent and batch-dependent matrix bias between CSS and native serum. Standard addition comparison revealed matrix factors of 1.005 to 1.611 for 5-methyltetrahydrofolate (5-mTHF), 0.607 to 1.000 for folic acid (FA), and 0.902 to 1.008 for folinic acid, indicating a higher analyte-to-internal-standard normalized response in native serum relative to CSS for 5-mTHF, a lower normalized response for FA, and minimal matrix-dependent response difference for folinic acid. To address this bias, we developed a pooled-serum, batch-specific matrix-factor correction workflow based on standard addition. This approach estimates matrix factors in pooled native serum and applies them to CSS-calibrated concentrations generated under the same batch conditions, avoiding the sample-volume burden of conventional standard addition. Matrix-factor correction reduced validation bias for 5-mTHF and FA to within the predefined ± 15% criterion. In 64 clinical serum samples, correction changed classification relative to selected total folate reference decision limits (5-21 ng/mL) in ten samples (15.6%) for 5-mTHF and five samples (7.8%) for FA. These findings show that CSS-based stable-isotope-dilution LC-MS/MS can produce residual matrix-dependent bias despite normalization using stable-isotope-labeled internal standards (SIL-IS), and that pooled-serum standard addition provides a practical correction strategy.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers