Evaluation and correction of charcoal-stripped serum matrix bias in stable-isotope-dilution LC-MS/MS quantification of endogenous serum folates.
Shuangshuang Zhu, Yik Lam Cho, Ying Deng, Meng Wei, Chan Weng Huen, Chui Mae Wong, Joo Guan Yeo, Xueming Dong
School of Biological Science, Nanyang Technological University, Singapore, 637551, Singapore. · Translational Immunology Institute, SingHealth Duke-NUS Academic Medical Centre, Singapore, 169856, Singapore.
Key points
From the paper's conclusion- These findings show that CSS-based stable-isotope-dilution LC-MS/MS can produce residual matrix-dependent bias despite normalization using stable-isotope-labeled internal standards (SIL-IS), and that pooled-serum standard addition provides a practical correction strategy.
Abstract
Isotope-dilution LC-MS/MS is widely used for targeted quantification of endogenous metabolites, but surrogate-matrix calibration can remain biased when the surrogate matrix is noncommutable with native specimens. In this study, three endogenous serum folate vitamers were quantified by stable-isotope-dilution LC-MS/MS using charcoal-stripped serum (CSS) as a surrogate blank matrix. Post-extraction spiking and pooled-serum standard addition showed analyte-dependent and batch-dependent matrix bias between CSS and native serum. Standard addition comparison revealed matrix factors of 1.005 to 1.611 for 5-methyltetrahydrofolate (5-mTHF), 0.607 to 1.000 for folic acid (FA), and 0.902 to 1.008 for folinic acid, indicating a higher analyte-to-internal-standard normalized response in native serum relative to CSS for 5-mTHF, a lower normalized response for FA, and minimal matrix-dependent response difference for folinic acid. To address this bias, we developed a pooled-serum, batch-specific matrix-factor correction workflow based on standard addition. This approach estimates matrix factors in pooled native serum and applies them to CSS-calibrated concentrations generated under the same batch conditions, avoiding the sample-volume burden of conventional standard addition. Matrix-factor correction reduced validation bias for 5-mTHF and FA to within the predefined ± 15% criterion. In 64 clinical serum samples, correction changed classification relative to selected total folate reference decision limits (5-21 ng/mL) in ten samples (15.6%) for 5-mTHF and five samples (7.8%) for FA. These findings show that CSS-based stable-isotope-dilution LC-MS/MS can produce residual matrix-dependent bias despite normalization using stable-isotope-labeled internal standards (SIL-IS), and that pooled-serum standard addition provides a practical correction strategy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparability of serum folate concentration measured by a microbiological method, LC-MS/MS and immunoassay.Annals of clinical biochemistry · 2026
- Quantitative Integration of Targeted and Nontargeted Metabolomics Using Multi-Stable Isotope Chemical Tagging with UHPLC-QToF MS: Application to Myeloid Leukemia Metabolism.Analytical chemistry · 2026
- B-370 Toward the Standardization of Homocysteine Testing: A Metrological Framework Based on a Candidate Reference Method and Commutable Candidate Reference MaterialsClinical Chemistry · 2026
- Matrix effects in untargeted LC-MS metabolomics: From creation to compensation with post-column infusion of standardsJournal of Chromatography A · 2025
- Formic Acid Pretreatment Enhances Untargeted Serum and Plasma MetabolomicsAnalytical Chemistry · 2025
- Accurate determination of vitamin B12 in infant formula using isotope dilution-ultrahigh-performance liquid chromatography-tandem mass spectrometry.Analytical and bioanalytical chemistry · 2026