Formic Acid Pretreatment Enhances Untargeted Serum and Plasma Metabolomics
Tereza Kačerová, Elisabete Pires, A. J. Dixon, Rachel Williams, Isabelle Legge, Mia Hippisley, Abi G. Yates, A. David Smith,
University of Oxford
Key points
From the paper's conclusion- In summary, we demonstrate that FA pretreatment of plasma and serum, prior to protein precipitation, significantly improves sample reproducibility and detection sensitivity in untargeted RPLC-MS metabolomics.
Abstract
Untargeted metabolic profiling of plasma and serum by liquid chromatography-mass spectrometry (LC-MS) is becoming increasingly important in clinical and translational research; however, sample preparation protocols can have a significant impact on study outcomes, and there is currently a lack of standardized approaches. In this study we demonstrate that pretreatment of serum and plasma samples with 1% formic acid (FA, v/v) prior to acetonitrile (MeCN)-induced protein precipitation significantly enhances analytical performance in untargeted metabolomics using reversed-phase liquid chromatography (RPLC)-MS. We show an increase in sample preparation reproducibility and signal intensity across both positive and negative ionization modes. In two independent serum cohorts (OPTIMA and VITACOG), FA-based extraction improved multivariate modeling (orthogonal partial least-squares discriminant analysis, OPLS-DA), with consistently higher classification accuracy, sensitivity, and specificity, alongside reduced variability and increased fold-changes in discriminatory compound-features. We investigated factors potentially involved in the enhanced performance and observed outcomes consistent with the disruption of noncovalent protein-metabolite interactions and the stabilization of labile species. We found no correlation with either protein depletion or differential adduct formation. The results were also not attributable to lowering pH after metabolite extraction. In summary, we demonstrate that FA pretreatment of plasma and serum, prior to protein precipitation, significantly improves sample reproducibility and detection sensitivity in untargeted RPLC-MS metabolomics. This optimized sample preparation strategy offers clear advantages for clinical and translational metabolomics, with the potential to enhance biomarker discovery and metabolic phenotyping.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Investigating Variability in Metabolomics: A Comparative Study of Analytical Platforms and Blood Matrices Using HPLC-HRMSMolecules · 2026
- MSlineaR: An R Package Assessing Linear Behavior to Improve Quality Assurance and Statistical Robustness in Untargeted Metabolomics.Analytical chemistry · 2026
- Propionic Acid Outperforms Formic and Acetic Acid in MS Sensitivity for High-Flow Reversed-Phase LC-MS Bottom-Up ProteomicsAnalytical Chemistry · 2026
- Matrix effects in untargeted LC-MS metabolomics: From creation to compensation with post-column infusion of standardsJournal of Chromatography A · 2025
- A chromatographic approach to avoid the matrix effect caused by sodium acetate clusters in the liquid chromatography coupled with electrospray ionization mass spectrometry analysis of free fatty acids in human plasmaJournal of Chromatography A · 2026
- MetaboRx: R Shiny application for quality assessment and preprocessing of mass spectrometry-based metabolomics data.bioRxiv : the preprint server for biology · 2026