Implementation and preliminary evaluation of a practical digital workflow for routine laser-radiation field coincidence quality assurance on the CyberKnife M6.
Linzheng Hu, Yongkang Zhao, Chunling Zheng, Jiajun Cai, Yajun Jia, Yi Tao, Yikun Liu, Yin Zhang,
National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital and Shenzhen Hospital Chinese Academy of Medical Sciences and Peking Union Medical College Shenzhen China. · School of Computer Science and Artificial Intelligence Wuhan University of Technology Wuhan China.
Key points
From the paper's conclusion- Purpose To implement and preliminarily evaluate a practical digital acquisition‐and‐analysis workflow for routine laser‐radiation field coincidence quality assurance (QA) on the CyberKnife M6 by integrating multileaf collimator (MLC) QA tool‐based laser reference definition, QA StereoChecker (QASC)‐based external digital image acquisition, and automated analysis with CKQA, an in‐house CyberKnife Q
Abstract
PURPOSE: To implement and preliminarily evaluate a practical digital acquisition-and-analysis workflow for routine laser-radiation field coincidence quality assurance (QA) on the CyberKnife M6 by integrating multileaf collimator (MLC) QA tool-based laser reference definition, QA StereoChecker (QASC)-based external digital image acquisition, and automated analysis with CKQA, an in-house CyberKnife QA software, and to compare this workflow with clinically realistic film-based alternatives. METHODS: Methods a and b were film-based measurements performed at the perch position. Method a used manual laser marking on film, whereas method b used the central metal marker of the MLC QA tool to define the laser reference. Method c used QASC as an external digital image-acquisition platform in combination with the MLC QA tool, and method d used film on the same QASC-based setup as a control. All methods were tested with 10 fixed collimators (5-40 mm). Images were analyzed using ImageJ and CKQA. Mean deviation, standard deviation, root mean square error (RMSE), and pairwise two-sided exact Wilcoxon signed-rank tests were used for exploratory comparisons. RESULTS: All evaluated workflows achieved submillimeter mean deviations. Across both analysis approaches, the mean deviation ranked consistently in descending order: method a > method b > method d > method c. Under the present experimental conditions, method c yielded the lowest mean deviation and RMSE (0.116 mm and 0.155 mm with CKQA), whereas method a showed the largest deviations. Dataset-level exact Wilcoxon signed-rank tests showed that several comparisons involving method a were statistically significant, but statistically significant differences were not uniform across both ImageJ- and CKQA-based comparisons. For comparisons between analysis methods within each workflow, CKQA and ImageJ were not significantly different for methods a and b, whereas CKQA yielded significantly lower deviations than ImageJ for methods c and d (p < 0.05). CONCLUSIONS: Laser-radiation field coincidence testing is an established CyberKnife QA procedure. The present study implemented and preliminarily evaluated a practical digital acquisition-and-analysis workflow for this routine task. The workflow combining MLC QA tool-based laser reference definition, QASC-based external digital image acquisition, and automated CKQA analysis provided a standardized and less operator-dependent procedure under the present experimental conditions. Its main value lies in reducing manual marking, film handling, scanning, and subjective image-analysis steps rather than in introducing a new physical measurement principle. In centers where QASC is not available, the film-based workflow using the MLC QA tool (method b) may serve as a practical alternative.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- ROC-based evaluation of tight gamma criteria for detecting MLC and jaw positioning errors in brain SRS using SRS MapCHECK.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Site-specific performance and complexity-dependent risk assessment of patient-specific quality assurance using a 2D array detector in CyberKnife radiosurgery.Radiological physics and technology · 2026
- Beyond Gamma Analysis: An Accessibility-oriented Framework for Comparative Evaluation of Patient-specific Radiotherapy Quality Assurance.Journal of medical physics · 2026
- Assessment of Clinical Quality Metrics for IROC Houston SRS Phantom Performance EvaluationDigital Commons-TMC (Texas Medical Center) · 2026
- Development and evaluation of a water-profile-based calibration method for 2D ionization chamber arrays in medical linear accelerator quality assurance.Journal of applied clinical medical physics · 2026
- Characterization and Validation of MicroDiamond Detectors in Patient-specific Quality Assurance of Magnetic Resonance-guided Stereotactic Radiotherapy.Journal of medical physics · 2026