Development and evaluation of a water-profile-based calibration method for 2D ionization chamber arrays in medical linear accelerator quality assurance.
Kihong Pak, Sung-Woo Kim, Byungchul Cho, Jungwon Kwak, Seonghoon Jeong, Kyeongyun Park, Jisoo Kim, Jonghan Park,
Department of Radiation Oncology, Asan Medical Center, Songpa-gu, Seoul, Republic of Korea. · Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine, Songpa-gu, Seoul, Republic of Korea.
Key points
From the paper's conclusion- For the tested 20 × 20 cm2 field on TrueBeam linacs with a CC13/Blue Phantom2 water-scanner reference, the proposed WP calibration method provided closer agreement with the water-scanner reference and a substantially shorter calibration time than the manufacturerprovided WF method. Machine-integrated calibration is recommended for routine multi-linac implementation, while energy-specific ACFs should be used whenever feasible to maximize accuracy.
Abstract
PURPOSE: To develop and evaluate a water-profile-based (WP) calibration method for two-dimensional (2D) ionization chamber arrays used in medical linear accelerator (linac) quality assurance (QA), and to compare its accuracy, efficiency, and cross-energy applicability with the manufacturer-provided wide-field (WF) calibration method. METHODS: Beam profiles were measured using an IC Profiler (ICP) and a 3D water scanner on three TrueBeam linacs (TB1-TB3) at five photon energies (4, 6, 10 MV flattened; 6, 10 MV flattening-filter-free [FFF]) with a 20 × 20 cm2 field size at a depth of 10 cm. A total of 240 profiles were acquired across eight independent measurement sessions. The WP method derived adjusted calibration factors (ACFs) by directly matching ICP detector responses to water-scanner reference profiles, and was evaluated under three strategies: (i) machine-specific (single linac reference), (ii) machine-integrated (reference pooled across three linacs), and (iii) cross-energy applicability across 20 inter-energy combinations. Performance metrics included per-detector relative differences (mean absolute error [MAE] and standard deviation [SD]) and beam profile characterization (symmetry and flatness). Paired t-tests (n = 30) compared the machine-integrated WP method with the WF method. RESULTS: The WF method yielded a per-detector MAE of 0.31%, with symmetry and flatness MAEs of 0.49% and 0.44%, respectively. The WP method significantly reduced these uncertainties to 0.15-0.16% (per-detector), 0.21-0.27% (symmetry), and 0.27-0.28% (flatness), with medium-to-large effect sizes (Cohen's d = 0.55-0.99; all p < 0.01). Cross-energy application of WP-derived ACFs yielded a per-detector MAE of 0.25%, higher than that of energy-specific WP calibration but lower than that of the WF method in this dataset. Total calibration time decreased from approximately 75 min (WF) to 10-15 min (WP). CONCLUSION: For the tested 20 × 20 cm2 field on TrueBeam linacs with a CC13/Blue Phantom2 water-scanner reference, the proposed WP calibration method provided closer agreement with the water-scanner reference and a substantially shorter calibration time than the manufacturerprovided WF method. Machine-integrated calibration is recommended for routine multi-linac implementation, while energy-specific ACFs should be used whenever feasible to maximize accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optimizing TrueBeam jaw calibrations to reduce positional errors at customizable gantry and collimator anglesJournal of Applied Clinical Medical Physics · 2026
- Experimental determination of beam quality correction factors in scanned proton beams using water calorimetryPhysics in Medicine and Biology · 2026
- Retrospective analysis of machine service logs and EPID-based linac QA to investigate the effects of machine maintenance and calibration events on beam characteristics: A multi-institutional study.Journal of applied clinical medical physics · 2026
- Are two readings as good as three for monthly linear accelerator output quality assurance?Journal of Applied Clinical Medical Physics · 2026
- Development of a pragmatic methodology for measurement of dose for KV CBCT in radiotherapy: A pilot study.Annals of the ICRP · 2026
- A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy.Biomedical physics & engineering express · 2026