Spectrometric characterization of x-ray attenuation in silicone-gel and saline breast implants under mammography and CT conditions.
Jérémie Lefèvre, Jade Kerfa, Mathis Bony, Johann Plagnard, Marion Catheline, Isabelle Fitton, Claire Von Ngoc Ty, Jean-Marc Bordy
Laboratoire National Henri Becquerel (LNE-LNHB), Université Paris-Saclay, CEA, List, Palaiseau, France. · Department of Radiology, AP-HP, Hôpital Européen Georges Pompidou, Paris, France.
Key points
From the paper's conclusion- These measurements supply energy-resolved attenuation data for silicone-gel and saline implants across the diagnostic range. PDMS and saline differ substantially at mammographic energies. Their monoenergetic attenuation converges in the Compton regime, but spectrum-weighted differences persist at 80 kV and should be accounted for in breast CT dosimetry. The results are consistent with recent phantom findings on AEC response and dose variations in the presence of silicone implants.
Abstract
BACKGROUND: Breast implants alter x-ray beam attenuation, but quantitative energy-resolved data for commercial prostheses are largely missing from the literature. Current dosimetric protocols lack specific correction factors for implant materials. PURPOSE: To measure the x-ray attenuation properties of silicone-gel (PDMS) and saline breast implants under mammography and computed tomography (CT) conditions. METHODS: We performed transmission spectrometry on six breast implants (three PDMS, three saline-filled) from a single manufacturer, using a cadmium telluride (CdTe) detector in narrow-beam geometry. Energy-dependent linear attenuation coefficients, effective attenuation coefficients ( μ eff ), half-value layers, and bulk densities were determined for mammography (25-49 kV) and CT (80-120 kV) beam qualities. Combined uncertainties were 3%-5%. RESULTS: Under mammographic conditions, PDMS attenuates more than saline: μ eff values are roughly 1.5-2 times higher, because of enhanced photoelectric absorption linked to a higher effective atomic number ( Z eff = 11.7 vs. 7.5). Saline implants behave much like water and soft breast tissues. The monoenergetic attenuation curves of the two materials converge above ∼ 60 keV as Compton scattering comes to dominate, but the spectrum-weighted effective attenuation still differs by ∼ 40% at a representative 80 kV Al CT tube potential (0.48 vs. 0.34 cm - 1 ), narrowing to ∼ 10% only with harder filtration and higher tube potential combined (120 kV AlCu). We provide log-log polynomial fits suitable for dosimetric codes. CONCLUSIONS: These measurements supply energy-resolved attenuation data for silicone-gel and saline implants across the diagnostic range. PDMS and saline differ substantially at mammographic energies. Their monoenergetic attenuation converges in the Compton regime, but spectrum-weighted differences persist at 80 kV and should be accounted for in breast CT dosimetry. The results are consistent with recent phantom findings on AEC response and dose variations in the presence of silicone implants.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and radiological characterization of low-cost breast tissue-equivalent phantoms made of epoxy resin-acetone for CT applications.Radiation and environmental biophysics · 2026
- Experimental determination of effective X-ray attenuation coefficients of 3D-printed materials under clinical mammography spectraFrontiers in Bioengineering and Biotechnology · 2026
- Normal and Pathological Incidental Breast Implant Findings on Chest Computed Tomography: What the Radiologist Needs to Know.Journal of breast imaging · 2026
- ABS and PETG as candidate materials for 3D-printed mammography phantoms: Physical and radiological characterization and assessment of TLD-indicated air kerma as a function of depth.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Computational study of gamma ray attenuation parameters of polymers using Phy-X/PSD and WinXcom ProgramsPhysics Open · 2026
- Radiation Shielding Properties of Polyethylene Reinforced with PbO and CdO: Experimental and Theoretical StudyNuclear Science and Engineering · 2026