Development and radiological characterization of low-cost breast tissue-equivalent phantoms made of epoxy resin-acetone for CT applications.

A Khallouqi, H Sekkat, W Allioui, A Slimani, A Halimi, O El Rhazouani

Faculty of Medicine and Pharmacy of Laâyoune, Ibn Zohr University, Laâyoune, Morocco. a.khallouqi@uhp.ac.ma. · Sciences and Engineering of Biomedicals, Biophysics and Health Laboratory, Higher Institute of Health Sciences, Hassan First University, Settat, 26000, Morocco.

Computational

Key points

From the paper's conclusion
  • All investigated materials provided a low-cost and tunable basis for breast phantom fabrication that warrants further evaluation for CT quality assurance and dosimetry applications.

Abstract

This study reports the fabrication and radiological characterization of three breast tissue-equivalent phantom materials for computed tomography (CT) dosimetry made of epoxy resin-acetone. Three samples were prepared at epoxy-to-acetone mass ratios of 80%/20% (S1), 60%/40% (S2), and 50%/50% (S3), and characterized through mass density measurements, NIST XCOM mass attenuation coefficient calculations, Phy-X/ZeXTRa effective atomic number (Zeff) determinations, and CT Hounsfield Unit (HU) measurements on a clinical Hitachi Supria 16 scanner at 80-140 kVp, benchmarked against a certified breast phantom insert from Computerized Imaging Reference Systems, Inc. (CIRS, Norfolk, VA, USA ) and International Commission on Radiological Protection (ICRP) Publication 44 reference data. Measured densities of 1.088, 0.994, and 0.954 g/cm³ for S1, S2, and S3, respectively, fall within the documented range of human breast tissue densities. In the Compton-dominated regime governing clinical CT, all three composites achieved close mass-attenuation agreement with ICRU-44 breast tissue, with absolute percentage difference (APD) values of 0.11%, 0.28%, and 0.44% at 80 keV. S3 provided the best overall µ/ρ agreement (mean APD = 7.15%, 20-150 keV) and Zeff equivalence at the Compton plateau (APD = 0.20%). Furthermore, S3 achieved the closest CT agreement, with a mean ΔHU of 0.2 HU against the CIRS ATOM reference at 100 kVp on the single scanner used here. Below approximately 60 keV, all three composites over-attenuate relative to breast tissue because of the residual chlorine of the epoxy matrix, so that the range of applicability supported by the present results is the Compton-dominated energy regime (≥ 60 keV) of routine CT. All investigated materials provided a low-cost and tunable basis for breast phantom fabrication that warrants further evaluation for CT quality assurance and dosimetry applications.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers