Do all ultra-congruent knees behave the same? A biomechanical insight into stability and kinematics.

Bernardo Innocenti

BEAMS Department, École polytechnique de Bruxelles Université libre de Bruxelles Brussels Belgium.

Key points

From the paper's conclusion
  • Purpose This study investigated the similarities and differences in stability and tibiofemoral kinematics between two ultra‐congruent (UC) total knee arthroplasty designs.

Abstract

PURPOSE: This study investigated the similarities and differences in stability and tibiofemoral kinematics between two ultra-congruent (UC) total knee arthroplasty designs. Specifically, the Link SymphoKnee (LSK) UC design was compared with the Zimmer Biomet Persona (PRSN) UC design under standardised loading and motion conditions. It was hypothesised that the designs would exhibit comparable overall patterns of stability and tibiofemoral kinematics, while differing in the magnitude of individual biomechanical outcomes. METHODS: Both implants were analysed using a previously validated finite element modelling framework, with identical anatomical, alignment and soft-tissue parameters. Anteroposterior (AP) and internal-external (IE) stability were evaluated under ±100 N AP forces and ±6 Nm IE torques at 0°, 30°, 60°, 90° and 120° of flexion. Three conditions were investigated: without collateral ligament prestrain under 44 N axial load, with prestrain under 44 N, and with prestrain under 500 N. Tibiofemoral kinematics and contact mechanics were also evaluated during standardised gait simulation and a loaded squat task up to 120° of flexion. Comparisons were descriptive, reflecting the deterministic nature of the simulations. RESULTS: Both designs showed increasing stability ranges from extension to mid-flexion, but the magnitudes differed between implants and loading conditions. Without collateral ligament prestrain, the maximum between-design differences in total AP and IE ranges were 7.40 mm and 9.06°, respectively. PRSN exhibited smaller AP and IE ranges at 0°, whereas LSK exhibited smaller ranges at 30°-120°. Increasing axial load with prestrain reduced both ranges in both designs. During standardised gait simulation, AP translation ranges were 9.9 mm for PRSN and 7.2 mm for LSK, whereas IE rotation ranges were 7.6° and 7.5°, respectively. Gait and squat simulations also revealed differences in kinematic trajectories and compartmental contact mechanics. CONCLUSION: The investigated UC configurations exhibited shared biomechanical response patterns alongside differences that depended on loading condition, flexion angle and outcome. LEVEL OF EVIDENCE: N/A, preclinical computational biomechanical study.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers