Effect of Joint Loadings, Ligament Tensions, and Total Knee Designs on Kinematics During Flexion.
Nicholas Parody, Joshua C Rozell, Peter S Walker
Department of Orthopedic Surgery, NYU Grossman School of Medicine, NYU Langone Health, 301 East 17th Street, NEW YORK, NY 10017. · Department of Orthopedic Surgery, NYU Grossman School of Medicine, NYU Langone Health, 301 East 17th Street, NEW YORK, NY 10017. Electronic address: ptrswlkr@aol.com.
Key points
From the paper's conclusion- The kinematics under low-load surgical conditions generally did not match the motion under functional conditions, except for the REP design, which produced similar motion for surgical low loads and functional high loads. For the REP design, the motion was controlled mainly by the contoured shapes of the bearing surfaces.
Abstract
BACKGROUND: Reproducing anatomic kinematics during flexion in a total knee arthroplasty (TKA) requires a stable medial side and progressive posterior displacement of the lateral femoral condyle. Under surgical conditions, however, the forces across the knee are relatively small, and it is not easy to predict functional kinematics where loads are much higher. The aims of this study were to determine the relative effects of the ligament tensions and the TKA itself on the kinematics that occur under low and high load conditions during a full range of flexion. METHODS: A crouching machine was developed with simulated femur, tibia, muscles, and collateral ligaments. There were four non-cruciate total knee arthroplasties that were designed: ultracongruent (UC), medially congruent (MC), medial pivot (MP), and replica intercondylar (REP), with guiding surfaces. There were two loading conditions: low load to simulate surgical conditions and high load for functional conditions. There were three collateral ligament tensions: normal, 0.5x normal, and 1.5x normal. The motion of the femur on the tibia was measured from 12 to 130° flexion. The transverse femoral axis projected on the tibia demonstrated displacements of the lateral and medial femoral condyles, as well as axial rotation. Representative pilot statistical tests were carried out for the MC and REP designs, with P = 0.1. RESULTS: Under low load, all designs showed small medial displacements, but with some lateral displacements and axial rotations. Under high load, the UC, MC, and MP showed almost parallel motion with small anterior-posterior (AP) displacements and rotations. The REP showed stable medial and more pronounced lateral displacements for both low and high loads. For all four designs, the behaviors were similar for the three ligament tension conditions. There was a statistically significant difference between rotations and lateral displacements of the MC and REP under low and high load. CONCLUSIONS: The kinematics under low-load surgical conditions generally did not match the motion under functional conditions, except for the REP design, which produced similar motion for surgical low loads and functional high loads. For the REP design, the motion was controlled mainly by the contoured shapes of the bearing surfaces.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Do all ultra-congruent knees behave the same? A biomechanical insight into stability and kinematics.Journal of experimental orthopaedics · 2026
- Lateral-stabilized total knee arthroplasty enables design-specific kinematic behavior consistent with key physiological characteristics during flexion.Frontiers in bioengineering and biotechnology · 2026
- Intraoperative medial compartment gaps were smaller than lateral compartment gaps across assessed flexion angles during restricted kinematic alignment total knee arthroplasty: a quantitative assessment using an independent tensor.The Knee · 2026
- Tibiofemoral and patellofemoral kinematics after kinematic versus mechanical alignment in varus TKA-a dynamic knee rig study.Archives of orthopaedic and trauma surgery · 2026
- Kinematic alignment reduces tibial internal rotation and maintains a larger retropatellar contact area than mechanical alignment in early flexion in valgus knees.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2026
- Exposing implausible knee loading: 3D load-balancing reveals limitations in traditional two-step musculoskeletal modeling.Journal of biomechanics · 2026