A-342 Method Comparison of Blood Gas Total Hemoglobin and Automated CBC Hemoglobin Measurements in a Neonatal Intensive Care Unit
Oytun Portakal, M U H A R R E M BAYRAM, S A M I L BINICI, T O L G A CELIK, F A T M A ARICIOGLU, B E D R I B E R A EKEN, Incilay Lay, G U N N U R DIKMEN
Key points
From the paper's conclusion- A total of 105 paired Hb measurements obtained from a blood gas analyzer in NICU and an automated CBC analyzer in central laboratory were retrospectively analyzed.
Abstract
Background Rapid hemoglobin (Hb) assessment using blood gas analyzers is common in neonatal intensive care units (NICUs). However, differences in analytical principles between co-oximetry-based total hemoglobin (tHb) and photometric CBC hemoglobin measurements may lead to clinically relevant discrepancies. Methods This study was conducted in Hacettepe University Children Hospital and The Central Laboratory. A total of 105 paired Hb measurements obtained from a blood gas analyzer in NICU and an automated CBC analyzer in central laboratory were retrospectively analyzed. Method comparison included independent t-tests, Passing–Bablok regression, Lin’s concordance correlation coefficient (CCC), and Bland–Altman analysis according to CLSI EP09-A3 recommendations. Results Blood gas Hb values were found to be higher with a mean positive bias of +0.72 g/dL. (p =0.027) Bland–Altman limits of agreement ranged from -1.29 to +2.72 g/dL. Passing–Bablok regression yielded a slope of 1.016 and an intercept of 0.43 g/dL indicating there is a small constant bias. Concordance was good (CCC=0.867). Conclusion Our results shows that although blood gas and CBC hemoglobin measurements demonstrate good correlation and concordance, a consistent positive bias exists. These two methods should not be considered fully interchangeable in NICU patients, in particular when strict biological variation–based quality specifications are applied. Method-specific interpretation and consistent monitoring using the same measuring platform are recommended.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of Blood Glucose Measurement Methods at Low Concentrations in Neonates at Risk for Hypoglycemia.The journal of applied laboratory medicine · 2026
- Blood Glucose Monitoring in the Neonatal Period: A Study on Evaluation Using the RAPIDPoint 500/500e.Pediatrics international : official journal of the Japan Pediatric Society · 2026
- Evaluation of Different Methods of Haemoglobin Estimation Among Malaria-Infected Patients.Journal of clinical laboratory analysis · 2026
- Validation of Non-Invasive Hemoglobin Measurement and Its Utility for Anemia Screening in Pregnancy: A Cross-Sectional Diagnostic Accuracy Study.Diagnostics (Basel, Switzerland) · 2026
- Research on Intraoperative Non-Invasive Hemoglobin Detection Technology Based on Four-Wavelength PhotoplethysmographyBiosensors · 2026
- Accuracy of minimally invasive hemoglobin monitoring in the perioperative period: transcutaneous hemoglobin monitoring and point-of-care testing devices: prospective observational study.Journal of clinical monitoring and computing · 2026