Snare-Assisted Retrieval of a Mechanically Locked-Open TEER Device: A Novel Bailout Technique.
Dhamodaran Kaliyamoorthy, Kannan Murugesan, Aishwarya Mahesh Kumar, Krishnaswamy Chandrasekaran
Apollo Hospitals, Chennai, India.
Key points
From the paper's conclusion- We report a novel bailout snare technique that enabled safe retrieval without the need for a second transseptal puncture.
Abstract
Transcatheter edge-to-edge repair (TEER) is an established intervention for patients having severe mitral regurgitation (MR) with prohibitive surgical risk. Complications during deployment related to device malfunction are uncommon but necessitate prompt decision-making and management. A 54-year-old gentleman with dilated cardiomyopathy, severe left ventricular dysfunction, and severe MR underwent TEER using the MyClip system (Meril Life Sciences Pvt. Ltd., Vapi, Gujarat, India). Initial deployment across the A2-P2 scallops of mitral leaflets achieved partial MR reduction. Significant residual MR required a second clip for optimal leaflet coaptation. While deploying the second clip, it failed to close because of a mechanical locking failure. So, we everted the clip, pulled back to LA, and it stayed in open position and failed to close. We report a novel bailout snare technique that enabled safe retrieval without the need for a second transseptal puncture. The ability to close the clip arms prior to withdrawal minimized the risk of tearing the atrial septum and facilitated the successful completion of the TEER procedure. Another clip was deployed successfully, resulting in significant MR reduction. This case demonstrates a rare complication during deployment, a malfunction of the device, and a novel strategy for safely bailing out a challenging intra-procedural complication.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Guide-sheath-assisted bailout technique for intra-procedural single-leaflet device attachment using the PASCAL system: a case report.European heart journal. Case reports · 2026
- Intraprocedural Embolization of a Tricuspid Edge-to-Edge Repair Device and Its Percutaneous Retrieval: A Case Report.Reports (MDPI) · 2026
- Successful Percutaneous Retrieval of an Embolized MitraClip From the Right Subclavian Artery.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Vascular Plug Closure of Delayed Leaflet Perforation Post-TEER in Patient With Multiple Implantable Devices.JACC. Case reports · 2026
- Clinical Failure After Mitral Transcatheter Edge-to-Edge Repair Mechanisms, Outcomes, and Management.Structural heart : the journal of the Heart Team · 2026
- Native Leaflet Interference After Transcatheter Mitral Valve-in-Ring: Snare-Assisted Valve-in-Valve-in-Ring Bailout.JACC. Case reports · 2026