Intraprocedural Embolization of a Tricuspid Edge-to-Edge Repair Device and Its Percutaneous Retrieval: A Case Report.
Andrija Matetic, Frane Runjić, Ivona Mustapić, Darija Baković Kramarić
Department of Cardiology, University Hospital of Split, Šoltanska 1, 21000 Split, Croatia.
Key points
From the paper's conclusion- Single-leaflet device attachment (SLDA) is the commonest device-related complication, but complete intraprocedural detachment with embolization is rare, and percutaneous retrieval of an embolized tricuspid device has been reported only exceptionally.
Abstract
Background and Clinical Significance: Tricuspid transcatheter edge-to-edge repair (T-TEER) is an established therapy for severe tricuspid regurgitation (TR) in selected patients. Single-leaflet device attachment (SLDA) is the commonest device-related complication, but complete intraprocedural detachment with embolization is rare, and percutaneous retrieval of an embolized tricuspid device has been reported only exceptionally. Case Presentation: An inoperable elderly patient with previous transcatheter aortic valve implantation (TAVI) and torrential functional TR underwent T-TEER. After implantation of the first clip at the anteroseptal commissure with apparently adequate leaflet capture, anterior SLDA was identified and progressed to complete detachment and right atrial embolization of the device. Despite this, the T-TEER was completed by successful implantation of the second clip, leaving only mild residual central TR. In the same session, the embolized device was retrieved percutaneously from the femoral venous system using a 15 mm gooseneck snare advanced through a 22 Fr sheath, with a safety wire and en-bloc removal, without forcing ensheathing and without venous injury. The patient was discharged on post-procedure day 2 with durable TR reduction and clinical improvement. Conclusions: Complete leaflet detachment is a possible complication of T-TEER. Confirming deep, symmetric insertion in multiple echocardiographic windows before every release is essential. When embolization occurs, knowledge of device dimensions and immediate availability of a complete snare kit is important. Although this approach may be considered in selected patients, particularly those at high surgical risk, the available evidence is limited to isolated case reports and does not allow general conclusions on its safety or efficacy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Successful Percutaneous Retrieval of an Embolized MitraClip From the Right Subclavian Artery.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Snare-Assisted Retrieval of a Mechanically Locked-Open TEER Device: A Novel Bailout Technique.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Device-Sparing Posteroseptal Trans-Tricuspid Lead Implantation After Tricuspid TEER: A Three-Case Device Rounds Report.Pacing and clinical electrophysiology : PACE · 2026
- Salvaging the forgotten valve: T-TEER as the new frontier in torrential primary TR.Frontiers in cardiovascular medicine · 2026
- Tricuspid Valve Transcatheter Edge-to-Edge Repair in Patients With Cardiac Implantable Electronic DevicesJACC: Cardiovascular Interventions · 2025
- Guide-sheath-assisted bailout technique for intra-procedural single-leaflet device attachment using the PASCAL system: a case report.European heart journal. Case reports · 2026