Beyond the Valve: A Tailored Distal Snaring Strategy for Complex Arterial and Venous Nose Cone Entrapment During Transfemoral TAVI and TricValve Implantation.
Adrien Jossart, Ignacio Amat Santos, Sara Blasco Turrion, Marouane Boukhris, Jacques Auslender, Giuseppe Colletti, Raphael Soetens, Mihai Cocoi,
Cardiovascular Jolimont Hospital, La Louvière, Belgium. · Division of Cardiology, Hospital Clínico Universitario de Valladolid, Valladolid, Spain.
Key points
From the paper's conclusion- The second occurred during heterotopic caval valve implantation (TricValve system), in which interference from a prominent Eustachian valve was suspected to have contributed to venous entrapment and dislodgement of the delivery system nose cone.
Abstract
Nose cone entrapment is a rare but potentially serious complication of structural heart interventions. Most cases have been reported during transcatheter aortic valve implantation (TAVI), typically related to heavy left ventricular outflow tract (LVOT) calcification, severe annular angulation, or under-expansion of the valve frame. Entrapment within vascular structures is much less common. We describe two unusual cases. The first involved femoral artery entrapment of the Evolut PRO+ nose cone due to a ProGlide suture knot, managed with a distal snaring technique but complicated by vascular injury requiring covered stent implantation. The second occurred during heterotopic caval valve implantation (TricValve system), in which interference from a prominent Eustachian valve was suspected to have contributed to venous entrapment and dislodgement of the delivery system nose cone. This was successfully retrieved en bloc with an EN Snare without vascular complications. These cases highlight that nose cone entrapment is not limited to valvular structures and may also occur in arterial or venous vasculature, with different mechanisms and bailout strategies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pigtail catheter entrapment during post-dilation causing Sapien-3 Ultra RESILIA valve damage with severe intravalvular aortic regurgitation: a case reportEuropean Heart Journal - Case Reports · 2026
- Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Use of EN Snare Device for Successful Retrieval of a Pigtail Catheter Straightener after Iliac Artery Stent Placement: A Case Report.Annals of vascular diseases · 2026
- Suture-mediated vascular closure for iatrogenic femoral artery pseudoaneurysms: a multicenter experience.Journal of vascular and interventional radiology : JVIR · 2026
- Snare-Assisted Transseptal Antegrade Transcatheter Aortic Valve Replacement: Turning Back the Clock.JACC. Case reports · 2026
- Gutter Rotation Technique for Percutaneous Ipsilateral Retrieval of a Ruptured SAPIEN-3 Balloon.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026