Pigtail catheter entrapment during post-dilation causing Sapien-3 Ultra RESILIA valve damage with severe intravalvular aortic regurgitation: a case report
Hideo Tokuyama, Hiroyuki Ikezaki, Naoki Sato
Key points
From the paper's conclusion- Discussion This case illustrates pigtail catheter entrapment during post-dilatation as a preventable cause of acute prosthetic leaflet damage and severe intravalvular regurgitation.
Abstract
Background Acute prosthetic leaflet dysfunction after transcatheter aortic valve implantation (TAVI) is rare but potentially life-threatening. Catheter-induced leaflet trauma during post-dilatation is less well recognised. Case summary An 88-year-old woman with severe symptomatic aortic stenosis and high-risk aortic root anatomy—including a narrow sinus of Valsalva, low left coronary artery (LCA) height, and a small, calcified sinotubular junction—underwent elective TAVI with a 23-mm Sapien 3 Ultra RESILIA valve deployed with a 2-mL underfill under prophylactic LCA protection. Initial transthoracic echocardiography and aortography showed trivial paravalvular leak and preserved LCA flow. During post-dilatation, the distal balloon tip and Safari guidewire snagged the pigtail catheter, pulling it across the prosthesis and into the left ventricle. Unrecognised in a crowded fluoroscopic field, the catheter became trapped between the expanding balloon and the valve frame, causing focal leaflet injury and severe intravalvular aortic regurgitation with haemodynamic collapse. Emergent valve-in-valve implantation with a 20-mm Sapien 3 Ultra RESILIA under bilateral coronary protection resolved the regurgitation, but displaced leaflet tissue caused LCA ostial compromise requiring chimney stenting. The patient recovered uneventfully and remained stable at 30-day follow-up. Discussion This case illustrates pigtail catheter entrapment during post-dilatation as a preventable cause of acute prosthetic leaflet damage and severe intravalvular regurgitation. It underscores the need for meticulous catheter management, critical reassessment of post-dilatation in older patients with trivial leak, and individualised valve platform and size selection in small, anatomically constrained aortic roots.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ventricularis delamination causing acute left main obstruction during TAVI: a case report.European heart journal. Case reports · 2026
- Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Electrosurgical Leaflet Modification to Prevent Coronary Obstruction During Transcatheter Aortic Valve Replacement in Failing Native and Bioprosthetic ValvesJACC: Cardiovascular Interventions · 2025
- Prosthesis-to-Prosthesis Balloon Interaction During TAVR Causing Aorto-Mitral Curtain Injury and Left Ventricular Pseudoaneurysm.JACC. Case reports · 2026
- Valve-in-Valve Transcatheter Aortic Valve Replacement Using the TRITON Technique: A Case Report.Journal of the Society for Cardiovascular Angiography & Interventions · 2026
- Severe paravalvular leakage caused by delayed dislocation of valve prosthesis after transcatheter aortic valve replacement for rheumatic aortic valve disease: A case report.Medicine · 2026