Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR.

Ryotaro Iwata, Daisuke Fukamachi, Naoya Matsumoto, Yasuo Okumura

Division of Cardiology, Department of Medicine, Nihon University Hospital, Tokyo, Japan. · Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.

Observational

Key points

From the paper's conclusion
  • This case suggests that rupture configuration may influence balloon collapse and retrievability, although the proposed relationship remains hypothesis-generating.

Abstract

Deployment-balloon rupture during balloon-expandable transcatheter aortic valve replacement can impede device retrieval. An octogenarian woman underwent transfemoral implantation of a 23-mm SAPIEN 3 Ultra RESILIA valve. The balloon ruptured at approximately the nominal inflation volume adjacent to focal sinotubular-junction calcification. Immediate aspiration produced complete collapse without residual contrast or abnormal balloon geometry, permitting resistance-free retrieval into the 14-F eSheath. Inspection demonstrated a longitudinal tear involving nearly the entire balloon length without circumferential separation or missing fragments. Tear orientation was identifiable only after retrieval; during the procedure, residual contrast, balloon geometry, response to aspiration, and withdrawal resistance guided management. This case suggests that rupture configuration may influence balloon collapse and retrievability, although the proposed relationship remains hypothesis-generating.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers