Longitudinal Deployment Balloon Rupture With Uncomplicated Retrieval During TAVR.
Ryotaro Iwata, Daisuke Fukamachi, Naoya Matsumoto, Yasuo Okumura
Division of Cardiology, Department of Medicine, Nihon University Hospital, Tokyo, Japan. · Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
Key points
From the paper's conclusion- This case suggests that rupture configuration may influence balloon collapse and retrievability, although the proposed relationship remains hypothesis-generating.
Abstract
Deployment-balloon rupture during balloon-expandable transcatheter aortic valve replacement can impede device retrieval. An octogenarian woman underwent transfemoral implantation of a 23-mm SAPIEN 3 Ultra RESILIA valve. The balloon ruptured at approximately the nominal inflation volume adjacent to focal sinotubular-junction calcification. Immediate aspiration produced complete collapse without residual contrast or abnormal balloon geometry, permitting resistance-free retrieval into the 14-F eSheath. Inspection demonstrated a longitudinal tear involving nearly the entire balloon length without circumferential separation or missing fragments. Tear orientation was identifiable only after retrieval; during the procedure, residual contrast, balloon geometry, response to aspiration, and withdrawal resistance guided management. This case suggests that rupture configuration may influence balloon collapse and retrievability, although the proposed relationship remains hypothesis-generating.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Gutter Rotation Technique for Percutaneous Ipsilateral Retrieval of a Ruptured SAPIEN-3 Balloon.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Balloon Rupture During Balloon-Expandable TAVR: Insights Into the Role of Calcium Distribution and Lesion Preparation From Two Cases.Journal of the Society for Cardiovascular Angiography & Interventions · 2026
- Pigtail catheter entrapment during post-dilation causing Sapien-3 Ultra RESILIA valve damage with severe intravalvular aortic regurgitation: a case reportEuropean Heart Journal - Case Reports · 2026
- Bailout cardiac surgery in patients undergoing transcatheter aortic valve replacement: a comprehensive analysis of post-marketing safety reportsmedRxiv · 2026
- Surgical Explantation of a Balloon-Expandable Transcatheter Aortic Valve With Aortic Root Preservation for Prosthetic Valve Endocarditis: A Case Report.Cureus · 2026
- Clinically silent early proximal seal failure with suspected interval distal endograft displacement after hybrid zone 2 thoracic endovascular aortic repair in severe arch tortuosity: a case report with 1-year follow-up.Frontiers in medicine · 2026