From the paper's conclusion
  • Here, we present a material-encoded capillary biointerface that integrates microporous media, microfluidic geometry, and a hydrophilic graphene/poly(3,4-ethylenedioxythiophene):poly(styrenesulfonate)/Prussian Blue bioelectronic interface into a unified, self-regulated architecture.

Abstract

Bioelectronic sensing in complex biological fluids is often constrained not by molecular recognition, but by poor coordination of transport, separation, and interfacial interactions at the material level. Here, we present a material-encoded capillary biointerface that integrates microporous media, microfluidic geometry, and a hydrophilic graphene/poly(3,4-ethylenedioxythiophene):poly(styrenesulfonate)/Prussian Blue bioelectronic interface into a unified, self-regulated architecture. Capillary forces generated within a microfiber network are redistributed through vent-regulated pathways and stabilized at a hydrophilic mixed ionic-electronic interface, enabling controlled plasma extraction, transport, and analyte delivery without external actuation. System performance is governed by the balance between capillary driving pressure and distributed hydrodynamic resistance, defining an operating regime that simultaneously achieves >99% cellular exclusion, ∼80-93% protein recovery, and stable interfacial wetting. Within this regime, undiluted whole blood is autonomously converted into a reproducible electrochemical signal. Using glial fibrillary acidic protein (GFAP) as a model biomarker, the platform enables direct sample-to-answer detection within ∼9 min, spanning 0.02 pg mL-1 to 200 ng mL-1 with an ultralow detection limit (∼0.01 fg mL-1) and reduced pre-analytical loss relative to centrifugation. These results establish that analytical performance is governed primarily by pre-analytical transport and interfacial regulation and introduce a generalizable nano/micro-engineered framework for autonomous, scalable point-of-care biointerfaces.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

MEC-BioSense: Material-Encoded Capillary Biointerface for Autonomous Whole-Blood Biomarker Sensing — Science Skim