A thread-based microfluidic electrochemical platform for whole-blood protein sensing using endogenous uric acid as a redox mediator.
Toru Nohgi, Sirawit Sornrak, Jixin Shang, Huizi Zhang, Ling Ding, Hiroshi Kawahira, Jun Kameoka
Graduate School of Information, Production and Systems, Waseda University, Kitakyushu, Fukuoka 808-0135, Japan. jkameoka@waseda.jp. · Medical Simulation Center, Jichi Medical University, Tochigi 329-0498, Japan.
Key points
From the paper's conclusion- Here, we present a thread-based microfluidic electrochemical platform for protein-related sensing in whole blood without external redox reagents by exploiting endogenous uric acid (UA) as a redox mediator.
Abstract
Electrochemical protein sensing in whole blood remains challenging because many existing assays rely on external redox probes, washing steps, and off-chip pretreatment such as plasma separation. Here, we present a thread-based microfluidic electrochemical platform for protein-related sensing in whole blood without external redox reagents by exploiting endogenous uric acid (UA) as a redox mediator. The device integrates capillary-driven plasma separation and electrochemical detection in a compact format. Within the thread network, blood cells are retained through a Ca2+-induced aggregation mechanism, enabling plasma delivery to the electrode interface without centrifugation. Target recognition is established through a blocking-type mechanism in which antigen-antibody complex formation suppresses the oxidation current of UA. Using C-reactive protein (CRP) as a model target, the platform showed concentration-dependent responses over a clinically relevant range. The limit of detection was 0.091 mg dL-1 in PBS and 0.106 mg dL-1 in the integrated blood-based device configuration. In a stepwise CRP spike-and-recovery experiment, recoveries of 103.9%, 99.6%, and 99.8% were obtained for three concentration levels, with all relative standard deviation (RSD) values below 1.20%. These results demonstrate excellent analytical accuracy and reproducibility of the proposed sensing platform. The inhibition-based signal was preserved across the physiologically relevant range of uric acid concentrations tested, suggesting that the sensing response is relatively insensitive to moderate variations in endogenous uric acid levels. In a preliminary porcine inflammation model, the CRP-related sensor response showed a delayed temporal association with WBC count, with a maximum cross-correlation of R = 0.85 at a lag of approximately 180 min. Together, these results validate the feasibility of reagent-free protein sensing directly from whole blood using endogenous electroactive molecules. This study introduces a simple sample-to-answer thread-based microfluidic platform and highlights a new strategy for decentralized biochemical analysis without external redox mediators.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- MEC-BioSense: Material-Encoded Capillary Biointerface for Autonomous Whole-Blood Biomarker Sensing.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026
- AuNPs/reduced graphene oxide biointerface with immobilized uricase for mediator-free electrochemical uric acid biosensing.Analytical and bioanalytical chemistry · 2026
- Wearable Microfluidic Electrochemical Platform Based on One-Pot Prussian Blue-Embedded Molecularly Imprinted Polymers for Simultaneous Detection of Sweat Uric Acid and Urea.ACS sensors · 2026
- Paper-based microfluidic electrochemical sensor for chloride ion and uric acid detection in sweatTalanta · 2025
- Real-time electrochemical protein monitoring using molecular pendulum sensorsbioRxiv (Cold Spring Harbor Laboratory) · 2026
- Rapid-Forming Conductive Antifouling Hydrogel Coating Enables Magnetically Guided Electrochemical POCT for Direct at‑Home Blood Testing.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026