Technical development and implementation of 3D‐QALAS on a 1.5T MR‐Linac for the brain: A prospective R‐IDEAL stage 0/1 technology development report
Lucas B. McCullum, Ashley Harrington, Natalie A. West, Grace Nehring, Clifton D. Fuller, Ergys Subashi, Brian A. Taylor, Ken‐Pin Hwang,
The University of Texas MD Anderson Cancer Center
Key points
From the paper's conclusion- Conclusions The technical feasibility of using 3D‐QALAS on the integrated 1.5T MR‐Linac was confirmed.
Abstract
BACKGROUND: Quantitative relaxometry of the brain on the integrated MRI / linear accelerator (MR-Linac) at high isotropic reconstructed resolution is currently limited due to prohibitively long scan times and limited field-of-views. PURPOSE: The purpose of this study was to assess the technical feasibility of the 3D-QALAS technique on the 1.5T MR-Linac which can acquire whole-brain 1 mm isotropic quantitative T1, T2, and PD maps along with multiple synthetic images in a 7-minute acquisition time. METHODS: A 1 mm isotropic 3D-QALAS acquisition was scanned in both phantoms and five healthy volunteers on the 1.5T Elekta Unity MR-Linac device with scan times around seven minutes. A test-retest protocol across five independent sessions for the phantom was conducted. The Lin's Concordance Correlation Coefficient (LCCC), repeatability, and reproducibility between measured and reference quantitative T1, T2, and PD values were determined in the phantom. Distortion was also studied. Vendor-provided reconstruction through SyMRI was performed to extract synthetic images and brain volume metric assessments in the healthy volunteers. RESULTS: The slope and LCCC between the measured and phantom reference values was 1.02 (1.00), 1.09 (0.90), and 0.99 (1.00) for T1, T2, and PD, respectively. Median distortion across the phantom remained below 2 mm. The repeatability and reproducibility coefficient-of-variation (CoV) was under 8% for all measured values. The measured brain volumes in the healthy volunteer was within expected age-adjusted reference values. CONCLUSIONS: The technical feasibility of using 3D-QALAS on the integrated 1.5T MR-Linac was confirmed. Applying this technique to the adaptive radiation therapy workflow of the brain will provide new opportunities to integrate quantitative imaging relaxometry biomarkers at 1 mm isotropic reconstructed resolution.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Simultaneous T1 and T2 Mapping in the Brain With QuantoRAGE: An MP2RAGE Variant Using T2-Prepared Inversion.Magnetic resonance in medicine · 2026
- Simultaneous T1, T2, R2*, PDFF, and triglyceride quantification in the head and neck using magnetic resonance fingerprinting on a 1.5T MR-Linac.Medical physics · 2026
- Omni-QALAS: Optimized Multiparametric Imaging for Simultaneous $T_{1}$, $T_{2}$ and Myelin Water Mapping.IEEE transactions on bio-medical engineering · 2026
- Quantitative mapping from conventional MRI using self-supervised physics-guided deep learning: Applications to a large-scale, clinically heterogeneous dataset.Medical image analysis · 2026
- Image quality evaluation of dual deep learning accelerated post-contrast Magnetization Prepared Rapid Gradient Echo (MPRAGE) imaging at 1.5 T and 3 T MRI.Acta radiologica (Stockholm, Sweden : 1987) · 2026
- Temporal evolution acquisition based arterial spin labeling (TEA-ASL) for accurate estimation of apparent T2 of perfusion signal in brain.Imaging neuroscience (Cambridge, Mass.) · 2026