Simultaneous T1 and T2 Mapping in the Brain With QuantoRAGE: An MP2RAGE Variant Using T2-Prepared Inversion.
Natalia Pato Montemayor, Tâm Johan Nguyên, José P Marques, Berk Can Açikgöz, Jessica A M Bastiaansen, Roland Kreis, Piotr Radojewski, Jonathan Stelter,
Swiss Innovation Hub, Siemens Healthineers International AG, Lausanne, Switzerland. · Laboratory of Computational Neuroscience, School of Computer and Communication Sciences and School of Life Sciences, EPFL, Lausanne, Switzerland.
Key points
From the paper's conclusion- QuantoRAGE enabled reproducible simultaneous T1 and T2 mapping at 3 T, providing an efficient framework for longitudinal and cross-sectional quantitative MRI studies and showed feasibility at 7 T.
Abstract
PURPOSE: To develop and validate a repeatable and reproducible approach, QuantoRAGE, for simultaneous whole-brain T1 and T2 mapping using adiabatic magnetization preparation. METHODS: QuantoRAGE is a 3D FLASH-based sequence using an adiabatic T2-prepared inversion followed by two readout blocks. Sequence repetitions with different preparation durations and inversion times provide T1- and T2-weighted images. Quantitative parameters are voxel-wise estimated by matching signal evolutions to a dictionary generated from extended phase graph simulations. To accelerate the method, the protocol was optimized using Cramér-Rao lower bound analysis, and a neural-network was trained on sequence-specific dictionaries for fast matching. The optimized protocol was evaluated in phantom and healthy volunteers at 3 T and compared against reference MP2RAGE T1 and multi-echo spin-echo T2 measurements. In vivo repeatability and reproducibility were quantified using scan-rescan experiments. Feasibility of QuantoRAGE at 7 T was tested in vivo. RESULTS: At 3 T, T1 and T2 phantom quantification agreed with nominal values (T1: r = 0.98, T2: r = 0.99). In eight healthy volunteers, high repeatability (bias ± SD, T1: 14.1 ± 30.2 ms, T2: 1.2 ± 1.7 ms) and reproducibility (T1: 18.9 ± 32.6 ms, T2: 1.1 ± 2.1 ms) were observed across brain regions, with T1 estimates comparable to reference methods (r = 0.96, slope = 0.74), and T2 underestimated (r = 0.79, slope = 0.44). Fast matching produced results consistent with dictionary-matching and significantly reduced parameter estimation time. At 7 T, the adiabatic preparation resulted in homogeneous T1 and T2 maps at 0.8-mm isotropic resolution. CONCLUSION: QuantoRAGE enabled reproducible simultaneous T1 and T2 mapping at 3 T, providing an efficient framework for longitudinal and cross-sectional quantitative MRI studies and showed feasibility at 7 T.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- High-Resolution Quantitative MRI Using Contrast-Domain Multiscale Energy (MuSE) RegularizationarXiv · 2026
- Quiet, rapid 3D multiparametric mapping using magnetization-prepared zero echo time MRIarXiv · 2026
- Quasi‐Steady‐State Equilibrium CEST Reconstruction Improves T 1 ‐Normalized Apparent Exchange‐Dependent Relaxation Analysis in Brain Tumor Patients at 3 TeslaNMR in Biomedicine · 2026
- Robust T1ρ , T2 , and T2 * mapping via spin-locked MOLED with synthetic data-driven deep learning reconstruction.Physics in medicine and biology · 2026
- Technical development and implementation of 3D‐QALAS on a 1.5T MR‐Linac for the brain: A prospective R‐IDEAL stage 0/1 technology development reportJournal of Applied Clinical Medical Physics · 2026
- Mapping R1 of fat, R1 of water, joint R2⁎, and proton density fat fraction at 3 T using the multi-echo MP2RAGE sequence.Magnetic resonance imaging · 2026